S-ROM*SLEEVE PRX ZT HA 18F-XXL
Report
- Report Number
- 1818910-2012-02798
- Event Type
- Injury
- Date Received
- January 31, 2012
- Date of Event
- January 6, 2012
- Report Date
- January 6, 2012
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 181891
- Product Code
- MEH
- PMA / PMN Number
- PK934412
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATIONS SINCE THEIR RELEASE FOR DISTRIBUTION. PER THE INITIAL REPORTING; PATIENT X-RAYS ARE NOT AVAILABLE FOR EXAMINATION. A REVIEW OF PROVIDED MEDICAL RECORDS CONFIRMS THE REPORTED LOOSENING. FROM THE INFORMATION PROVIDED IT APPEARS THE S-ROM SLEEVE COMPONENT WAS CEMENTED INTO PLACE. THE INSTRUCTION FOR USE (IFU) STATES THAT THE SLEEVE COMPONENTS ARE INDICATED FOR CEMENT-LESS USE. FROM THE INFORMATION AVAILABLE IT IS LIKELY THAT THIS IS UNJUSTIFIED WITH RESPECT TO THE DEVICE. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.
(B)(4).
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES FINDS NO OTHER REPORTS AGAINST THE PRODUCT AND LOT CODE COMBINATIONS SINCE THEIR RELEASE TO DISTRIBUTION. MEDICAL RECORDS WERE RECEIVED AND REVIEWED. THE INVESTIGATION CAN DRAW NO CONCLUSION WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED.DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER RELATED REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT CODE/LOT NUMBER COMBINATION SINCE RELEASE FOR DISTRIBUTION. BASED ON PREVIOUS INVESTIGATIONS, THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
UPDATE REC'D 05/02/2014 - PFS AND MEDICAL RECORDS RECEIVED. PART/LOT WAS PROVIDED. AFTER REVIEW OF THE MEDICAL RECORDS THEY INDICATED THE PATIENT HAD ALL IMPLANTS REMOVED DURING THE REVISION DUE TO A POSSIBLE LOW GRADE INFECTION. THE CUP IS NOW BEING ADDED TO THE COMPLAINT, BUT NOT REPORTED AS IT WAS IMPLANTED ABOUT 3 YEARS AGO. THE REVISION OPERATIVE NOTE DID CONFIRM A LOOSE STEM. THERE WAS NO MENTION OF METALLOSIS. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE EXISTING MDR DECISION.
PATIENT WAS REVISED TO ADDRESS HIP PAIN AND SROM STEM LOOSENING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | S-ROM*SLEEVE PRX ZT HA 18F-XXL | HIP FEMORAL STEM/SLEEVE | MEH | DEPUY ORTHOPAEDICS, INC. 181891 | 2704233 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Required Intervention |