FDA Adverse Event Malfunction Summary report: N

MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 24320398 · Received February 12, 2026

Report

Report Number
2032227-2026-125601
Event Type
Malfunction
Date Received
February 12, 2026
Date of Event
January 23, 2026
Report Date
May 4, 2026
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
PMA / PMN Number
P160017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE UNIT RECEIVED PUMP ERROR 38 DURING TESTING. AFTER FURTHER REVIEW, THE MOTOR WAS UNABLE TO TURN DUE TO A JAMMED GEARBOX. AFTER FURTHER REVIEW, THE MOTOR WAS UNABLE TO TURN DUE TO A JAMMED GEARBOX. UNABLE TO PERFORM THE DISPLACEMENT, REWIND, PRIME/SEATING, BASIC OCCLUSION, FORCE SENSOR DUE TO PUMP ERROR 38. THUMP SOFTWARE WAS UTILIZED AND DOWNLOADED TRACE/HISTORY FILES PROPERLY. IN FURTHER FULL REVIEW IN THE PUMP HISTORY FOUND MULTIPLE PUMP ERROR 38 ON (B)(6) 2026 10:01:37.000, (B)(6) 2026 10:10:35.000, (B)(6) 2026 10:47:06.000, (B)(6) 2026 10:54:53.00. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND NO COMPONENT OR MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY, BATTERY CONNECTOR, BATTERY TUBE, MOTOR, VIBRATOR DURING VISUAL INSPECTION. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE, STAINED KEYPAD OVERLAY. PUMP ERROR 38 DURING TESTING AND IN THE PUMP HISTORY CONFIRMED DUE TO A JAMMED GEARBOX. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED A PUMP ERROR 38(NO MOTOR MOVEMENT OR NEGLIGIBLE MOVEMENT. RETRIES TO FREE A STUCK MOTOR FAILED). THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1884. TROUBLESHOOTING WAS PARTIALLY PERFORMED; THE CUSTOMER WAS ABLE TO CLEAR THE ALARM AND COMPLETE THE INSULIN PUMP REWIND AND THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL BE REPLACED AND INSTRUCTED TO REVERT TO A BACKUP PLAN PER HEALTH CARE PROFESSIONAL INSTRUCTIONS AND PLACE PUMP IN STORAGE MODE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE CUSTOMER WILL DISCONTINUE USE OF THE INSULIN PUMP. MMT-1884 WAS REQUESTED AND THE CUSTOMER RESPONSE WAS THAT THE DEVICE WILL BE RETURNED.