STATSTRIP GLUCOSE HOSPITAL METER SYSTEM
Report
- Report Number
- 1219029-2026-00012
- Event Type
- Malfunction
- Date Received
- February 11, 2026
- Date of Event
- January 16, 2026
- Report Date
- February 11, 2026
- Manufacturer
- NOVA BIOMEDICAL CORP.
- Product Code
- PZI
- UDI-DI
- 00385480547901
- PMA / PMN Number
- K150281
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THIS ISSUE WAS ESCALATED TO THE NOVA CONNECTIVITY SUPPORT TEAM AND R&D. AFTER TROUBLESHOOTING THE TRANSMISSION PROBLEMS, AN ISSUE WITH THE CUSTOMER'S NETWORK SETTINGS WAS IDENTIFIED AND RESOLVED. THE METER SYSTEM WAS PERFORMING AS INTENDED. TECHNICAL SUPPORT CONFIRMED THAT ALL THEIR METERS WERE CONNECTING AND TRANSFERRING RESULTS PROPERLY AFTER THE CHANGES WERE MADE. NO FURTHER ACTION IS RECOMMENDED AT THIS TIME. NOVA BIOMEDICAL WILL CONTINUE TO MONITOR FOR THIS OR SIMILAR EVENTS.
CUSTOMER IS REPORTING THAT THEY'RE SEEING DELAYS IN RESULTS BEING SENT FROM THIER NOVA STATSTRIP 1.86 CONNECTIVITY METERS TO THEIR QML MIDDLEWARE. ALTHOUGH THERE WERE NO DIRECT ALLEGATIONS OF DELAYS IN TREATMENT DUE TO THIS ISSUE, THE POTENTIAL FOR SUCH DELAYS EXISTS FOR ISSUES OF THIS NATURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 379678 | STATSTRIP GLUCOSE HOSPITAL METER SYSTEM | PRESCRIPTION USE BLOOD GLUCOSE METER FOR NEAR-PATIENT TESTING | PZI | NOVA BIOMEDICAL CORP. | 1.86 | 00385480547901 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |