PINNACLE MTL INS NEUT36IDX58OD
Report
- Report Number
- 1818910-2012-02480
- Event Type
- Injury
- Date Received
- January 30, 2012
- Date of Event
- November 28, 2011
- Report Date
- November 28, 2011
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 1818910
- Product Code
- KWA
- PMA / PMN Number
- K003523
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FOUND ADDITIONAL COMPLAINTS AGAINST THE 2222087 AND 2259361 LOT CODES. PER WI-3430, REVISION C, A REVIEW OF THE DEVICE HISTORY RECORDS FOR THESE PRODUCTS IS NO LONGER REQUIRED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE REMAINING PRODUCT AND LOT COMBINATION. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
**UPDATE** 01/25/2012 - LITIGATION PAPERS WERE RECEIVED. LITIGATION ALLEGES AFTER THE SURGERIES, HE PATIENT EXPERIENCED SEVERE PAIN AND DISCOMFORT AND INFLAMMATION IN HIS LEFT AND RIGHT THIGHS AND GROIN. HE ALSO EXPERIENCED A POPPING AND SNAPPING SENSATION IN HIS HIP-JOINTS WHEN WALKING OR MOVING TO AND FROM A SITTING POSITION. DOI: (B)(6) 2006 THE METAL-ON-METAL DEVICES HAVE BEEN ADDED TO THE COMPLAINT AS THEY ARE THE SUBJECT OF THE LITIGATION. THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS FOUND NO RELATED MANUFACTURING DEVIATIONS OR RELATED ANOMALIES. A SEARCH OF THE COMPLAINT DATABASE FOUND NO PRIOR REPORTS FOR STEM AND METAL INSERT PART AND LOT NUMBER COMBINATIONS; ONE PRIOR REPORT FOR THE HEAD PART AND LOT NUMBER COMBINATION. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED
(B)(4).
PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PATIENT WAS REVISED TO ADDRESS STEM LOOSENING. **UPDATE** 01/25/2012 - LITIGATION PAPERS WERE RECEIVED. LITIGATION ALLEGES AFTER THE SURGERIES, HE PATIENT EXPERIENCED SEVERE PAIN AND DISCOMFORT AND INFLAMMATION IN HIS LEFT AND RIGHT THIGHS AND GROIN. HE ALSO EXPERIENCED A POPPING AND SNAPPING SENSATION IN HIS HIP-JOINTS WHEN WALKING OR MOVING TO AND FROM A SITTING POSITION.
UPDATE REC'D 10/10/2012 - PFS AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDIAL RECORDS IT CONFIRMED LOOSENING OF THE FEMORAL STEM. THERE IS NO NEW ADDITIONAL INFORMATION THAT WOULD AFFECT THE EXISTING MDR DECISION.
IN ADDITION TO WHAT WERE PREVIOUSLY ALLEGED, PPF ALLEGES FRACTURE (BONE). ADDED UNKNOWN HIP IMPLANT. ADDED REVISION HOSPITAL, REVISION SURGEON AND LAWYER IN THE ASSOCIATED CONTACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PINNACLE MTL INS NEUT36IDX58OD | PINNACLE HIP SYSTEM : HIP METAL ACETABULAR LINERS | KWA | DEPUY ORTHOPAEDICS, INC. 1818910 | 2222087 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Required Intervention |