FDA Adverse Event Injury Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 24313705 · Received February 11, 2026

Report

Report Number
3008642652-2026-01421
Event Type
Injury
Date Received
February 11, 2026
Date of Event
February 3, 2026
Report Date
February 11, 2026
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005005
PMA / PMN Number
P010030
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE MONITOR AND ELECTRODE BELT HAVE NOT BEEN RETURNED FOR EVALUATION. BASED ON THE DATA AVAILABLE AT THIS TIME, THERE IS NO INDICATION OF A DEVICE MALFUNCTION CAUSING OR CONTRIBUTING TO THE INAPPROPRIATE TREATMENT. INAPPROPRIATE DEFIBRILLATIONS ARE AN ANTICIPATED RISK ASSOCIATED WITH THE USE OF THE LIFEVEST. PATIENTS ARE INSTRUCTED THROUGH ALARMS, VOICE MESSAGES, IFU, AND TRAINING TO PRESS THE RESPONSE BUTTONS TO PREVENT AN INAPPROPRIATE DEFIBRILLATION. THE CURRENT COMMERCIAL INAPPROPRIATE DEFIBRILLATION RATE IS CONSISTENT WITH THE OBSERVED RATE DURING THE PIVOTAL CLINICAL TRIAL G960083 (0.69% PER PATIENT-MONTH WITH 90% CONFIDENCE). A SUMMARY OF THE SAFETY AND EFFECTIVENESS DATA (SSED), INCLUDING THE INAPPROPRIATE DEFIBRILLATION SAFETY OBJECTIVE SUPPORTING FDA'S APPROVAL OF THE LIFEVEST, CAN BE FOUND AT HTTP://WWW.ACCESSDATA.FDA.GOV/CDRH_DOCS/PDF/P010030B.PDF.

Description of Event or Problem · 0

A PATIENT RECEIVED AN INAPPROPRIATE SHOCK. SVT CONTRIBUTED TO THE FALSE DETECTION. THE RESPONSE BUTTONS WERE PRESSED AFTER TREATMENT. THE PATIENT DID NOT SEEK MEDICAL ATTENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
379865 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM 00855778005005

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown