FDA Adverse Event Injury Summary report: N

GLIDEWELL HT IMPLANT Ø4.3 X 10 MM

MDR report key: 24313512 · Received February 11, 2026

Report

Report Number
3011649314-2026-00213
Event Type
Injury
Date Received
February 11, 2026
Date of Event
December 18, 2025
Report Date
April 11, 2026
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K143353
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED FOR ANALYSIS. COMPLAINT HISTORY AND PRODUCT HISTORY RECORDS WERE REVIEWED, AND DOCUMENTATION INDICATED THE PRODUCT MET RELEASE CRITERIA. ROOT CAUSE HAS NOT BEEN IDENTIFIED. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Additional Manufacturer Narrative · 0

THE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DHR RESULTS: THE DHR WAS REVIEWED FOR GLIDEWELL HT IMPLANT LOT# 6250253 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. STOCK PRODUCT REVIEWED RESULTS: A REVIEW OF STOCK PRODUCT WAS PERFORMED FOR GLIDEWELL HT IMPLANT LOT# 6250253 AND FOUND NO ADDITIONAL PRODUCT IN STOCK. INVESTIGATION METHODS/RESULTS: THE DEVICE WAS RETURNED BUT NOT IN THE ORIGINAL PACKAGE. A DENTAL PROSTHETIC WAS ATTACHED TO THE IMPLANT. THE IMPLANT WAS VERIFIED TO BE A GLIDEWELL HT IMPLANT Ø4.3 X 10 MM (70-1189-IMP0010) USING THE RADIOGRAPHIC TEMPLATE (MKT-013579 REV 1 PK-4500230-112023). THERE WAS NO DEFECT OR NON-CONFORMITY OBSERVED AND THE THREADS WERE INTACT. MATTER WAS OBSERVED IN THE TREADING OF THE IMPLANT. THE COMPLAINT IS VERIFIED BASED ON THE RETURNED PART BUT CANNOT CONFIRM THE FAILURE MODE. THERE WAS NO EVIDENCE FOUND THAT INDICATED THAT THE REPORTED ISSUE WAS CAUSED BY THE DEVICE ITSELF. ROOT CAUSE: "LOSS OF OSSEOINTEGRATION" IS A COMMON COMPLAINT IN REGARD TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES. ALTHOUGH THE ROOT CAUSE FOR LOSS OF OSSEOINTEGRATION IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT, OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS. PER THE REPORTED INFORMATION, THE PATIENT HAS TYPE II BONE QUALITY. IFU-012631 REV 2 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE CONTRAINDICATIONS SECTION: "GLIDEWELL HT IMPLANTS SHOULD NOT BE PLACED IN PATIENTS DISCOVERED TO BE MEDICALLY UNFIT FOR THE INTENDED TREATMENT. PRIOR TO CLINICAL INTERVENTION, PROSPECTIVE PATIENTS MUST BE THOROUGHLY EVALUATED FOR ALL KNOWN RISK FACTORS AND CONDITIONS RELATED TO ORAL SURGICAL PROCEDURES AND SUBSEQUENT HEALING. CONTRAINDICATIONS INCLUDE BUT ARE NOT LIMITED TO: VASCULAR CONDITIONS, UNCONTROLLED DIABETES, CLOTTING DISORDERS, ANTICOAGULANT THERAPY, METABOLIC BONE DISEASE, CHEMOTHERAPY OR RADIATION THERAPY, CHRONIC PERIODONTAL INFLAMMATION, INSUFFICIENT SOFT TISSUE COVERAGE, METABOLIC OR SYSTEMIC DISORDERS ASSOCIATED WITH WOUND AND/OR BONE HEALING, USE OF PHARMACEUTICALS THAT INHIBIT OR ALTER NATURAL BONE REMODELING, ANY DISORDERS WHICH INHIBIT A PATIENT'S ABILITY TO MAINTAIN ADEQUATE DAILY ORAL HYGIENE, UNCONTROLLED PARAFUNCTIONAL HABITS, INSUFFICIENT HEIGHT AND/OR WIDTH OF BONE, AND INSUFFICIENT INTERARCH SPACE". IFU-012631 REV 2 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE SURGICAL PROCEDURES UNDER THE PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. FOR BEST RESULTS, PLEASE OBSERVE THE FOLLOWING PRECAUTIONS: ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL, COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." IFU-012631 REV 2 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "THE IMPLANT SITE SHOULD BE INSPECTED FOR ADEQUATE BONE BY RADIOGRAPHS, PALPATIONS AND VISUAL EXAMINATION. DETERMINE THE LOCATION OF NERVES AND OTHER VITAL STRUCTURES AND THEIR PROXIMITY TO THE IMPLANT SITE BEFORE ANY DRILLING TO AVOID POTENTIAL INJURY, SUCH AS PERMANENT NUMBNESS TO THE LOWER LIP AND CHIN." IFU-012631 REV 2 (GLIDEWELL HT IMPLANT SYSTEM IFU) CONTAINS THE FOLLOWING STATEMENT IN THE WARNINGS SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE, AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." THE MANUFACTURER INTERNAL REFERENCE NUMBER IS: (B)(4).

Description of Event or Problem · 0

ON (B)(6) 2026, DR.(B)(6) REPORTED THAT HIS 35-YEAR-OLD MALE PATIENT CAME INTO HIS ONA HEALTHY SMILE OFFICE AFTER (B)(6) MONTHS OF FINAL PROSTHESIS DELIVERY. THE PRIMARY IMPLANT PLACEMENT SURGERY AT TOOTH #19 WAS PERFORMED ON (B)(6) 2025, FOLLOWED BY PLACEMENT OF A SINGLE-TOOTH, CEMENT-RETAINED PROSTHETIC RESTORATION PERFORMED ON (B)(6) 2025. LOSS OF OSSEOINTEGRATION OF THE IMPLANT WAS IDENTIFIED ON (B)(6) 2025. IT WAS REPORTED THAT THE IMPLANT WAS REMOVED AND REPLACED. THE PATIENT DID NOT SUFFER PERMANENT INJURY OR REQUIRE ADDITIONAL MEDICAL OR SURGICAL PROCEDURES. THE PATIENT'S BONE QUALITY IS TYPE II WITH GOOD ORAL HYGIENE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373633 GLIDEWELL HT IMPLANT Ø4.3 X 10 MM GLIDEWELL HT IMPLANT SYSTEM DZE PRISMATIK DENTALCRAFT, INC. 70-1189-IMP0010 6250253

Patients

Seq Age Sex Outcome Treatment
1 35 YR Male Required Intervention