FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION¿ S8 SYSTEM

MDR report key: 24313505 · Received February 11, 2026

Report

Report Number
1723170-2026-00222
Event Type
Malfunction
Date Received
February 11, 2026
Date of Event
February 9, 2026
Report Date
February 24, 2026
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
UDI-DI
00763000272739
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9 735772: (H3,H6) NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H2: PLEASE SEE SECTION D9 FOR WHEN THE DEVICE WAS AVAILABLE FOR ANALYSIS. H2-H6: A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE INTERCONNECT CABLE BETWEEN MAIN CART AND CAMERA CART WAS REPLACED AND THE SYSTEM PERFORMED AS INTENDED. CODES B01, C13, AND D02 ARE APPLICABLE. H2-H6: THE INTERCONNECT CABLE, LOT: 190326, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE DAMAGE WAS VERIFIED. THE RETURNED INT ERCONNECT CABLE WAS TORN WITH EXPOSED WIRES. DESPITE THE DAMAGE, THE CABLE WAS TESTED AND WAS FUNCTIONAL. CODES B01, C02, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THERE WERE EXPOSED WIRES ON THE INTERCONNECT CABLE WHICH WAS LIKELY DUE TO WEAR AND TEAR. THERE WAS NO PATIENT PRESENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
373491 STEALTHSTATION¿ S8 SYSTEM NEUROLOGICAL STEREOTAXIC INSTRUMENT HAW MEDTRONIC NAVIGATION, INC. 9735665 00763000272739

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown "SEE H11....".