FDA Adverse Event Injury Summary report: N

TRINICA SCREW SELF-DRILLING VARIABLE 4.2X14

MDR report key: 2431041 · Received January 26, 2012

Report

Report Number
2184052-2012-00002
Event Type
Injury
Date Received
January 26, 2012
Date of Event
December 29, 2011
Report Date
December 30, 2011
Manufacturer
ZIMMER SPINE, INC.
Product Code
KWQ
PMA / PMN Number
K012305
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

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

Description of Event or Problem · 1

IT WAS REPORTED THE PT PRESENTED WITH NECK PAIN APPROXIMATELY 22 MONTHS POST-OP. X-RAYS SHOWED BROKEN SCREWS AT C3 AND C6. THE CONSTRUCT WAS REMOVED HOWEVER THE TWO BROKEN SCREW TIPS REMAIN IN THE VERTEBRAL BODY AT C3 AND C6. THE PT WAS REVISED WITH A DIFFERENT DEVICE. NO FURTHER PT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRINICA SCREW SELF-DRILLING VARIABLE 4.2X14 TRINICA SCREW SELF-DRILLING VARIABLE KWQ ZIMMER SPINE, INC. 07.00812.005 61189062

Patients

Seq Age Sex Outcome Treatment
1 71 YR Required Intervention