FDA Adverse Event Malfunction Summary report: N

MODULAR FLEXIBLE REAMER HEAD, Ø15.5MM

MDR report key: 24308866 · Received February 10, 2026

Report

Report Number
1220246-2026-00512
Event Type
Malfunction
Date Received
February 10, 2026
Date of Event
January 16, 2026
Report Date
March 13, 2026
Manufacturer
ARTHREX, INC.
Product Code
HSB
UDI-DI
00848665000986
PMA / PMN Number
K191598
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED UPON AVAILABILITY OF ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION: D9, G3, H3, H6. THE COMPLAINT ALLEGATION IS CONFIRMED. - ONE UNPACKAGED 0234-155 SERIAL/BATCH NUMBER 211146 WAS RECEIVED FOR INVESTIGATION. - FUNCTIONAL TESTING WAS NOT PERFORMED DUE TO THE DAMAGE TO THE DEVICE. - VISUAL INSPECTION NOTED THAT THE DEVICE IS SIGNIFICANTLY DAMAGED WHERE THE FEMORAL SHAFT ATTACHES. - THE MOST LIKELY CAUSE IS ATTRIBUTED TO MISUSE/USE ERROR DUE TO IMPROPER BONE PREP; MISALIGNED INSERTION; PRYING/LEVERAGING THE DEVICE DURING INSERTION.

Description of Event or Problem · 0

ON 17-JAN-2026, A FACILITY REPRESENTATIVE REPORTED VIA (B)(4) THAT A 0234-155 15.5 MM MODULAR FLEXIBLE REAMER HEAD BROKE DURING THE INTRAMEDULLARY REAMING OF A FEMORAL SHAFT. THE HIP NAIL PROCEDURE WAS COMPLETED SUCCESSFULLY; HOWEVER, THE BROKEN PIECE WAS NOT RETRIEVED FROM THE PATIENT. IT WAS REPORTED THAT THERE WAS NO OTHER PATIENT HARM. ADDITIONAL INFORMATION WAS RECEIVED ON 04-FEB-2026. THE REP ADVISED THAT THERE WERE TWO SMALL FRAGMENTS CREATED IN ADDITION TO THE LARGER REMAINING PORTION OF THE REAMER THAT WAS RETRIEVED, AND THOSE TWO FRAGMENTS REMAINED IN THE PATIENT. THE REP STATED THAT NO ATTEMPT WAS MADE TO RETRIEVE THE FRAGMENTS, AS THEY WERE MIDSHAFT AND DEEMED IMPOSSIBLE TO RETRIEVE. THERE WAS NO DELAY ENCOUNTERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
42483 MODULAR FLEXIBLE REAMER HEAD, Ø15.5MM REAMER HSB ARTHREX, INC. MODULAR FLEXIBLE REAMER HEAD, Ø15.5MM 211146 00848665000986

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown