I-STAT CHEM8+ CARTRIDGE
Report
- Report Number
- 2245578-2026-00027
- Event Type
- Malfunction
- Date Received
- February 10, 2026
- Date of Event
- January 14, 2026
- Report Date
- April 13, 2026
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- JPI
- PMA / PMN Number
- K183680
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 09-APR-2026. A REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING OF CHEM8+ LOT H25257 AND RETAINED AND RETURNED CARTRIDGE TESTING OF CHEM8+ CARTRIDGE LOT H25301 MET THE ACCEPTANCE CRITERIA PER Q04.01.003 REV. AP, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOTS H25257 AND H25301.
APOC INCIDENT # (B)(4) APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
NA.
ON (B)(6) 2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT IONIZED CALCIUM RESULT ON AN 8-MONTH-OLD FEMALE PEDIATRIC PATIENT IN THE ICU. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD: DATE: COLLECTED TESTED RESULTS: SAMPLE I-STAT, (B)(6) 2026 , 19:59 , 19:59 , K: 6.8 MMOL/L, GLUC: 166 MG/DL, TCO2: 14 MMOL/L, BUN: 19 MG/DL, ICA: 0.80 MMOL/L A I-STAT, (B)(6) 2026, 20:002 , 20:05 K: 6.9 MMOL/L, GLUC: 210 MG/DL, TCO2: 12 MMOL/L, BUN: 18 MG/DL, ICA: 1.39 MMOL/L B. ROCHE (B)(6) 2026 , 20:02 , 20:30 , K: 7.9 MMOL/L, GLUC: 283 MG/DL, TCO2: 7 MMOL/L, BUN: 14 MG/DL . C. I-STAT , (B)(6) 2026, 21:15 , 21:21 , K: 6.5 MMOL/L, GLUC: 304 MG/DL. CHLORIDE: 120 MMOL/L . D ROCHE (B)(6) 2026 , 21:15, 21:31, K: 9.1 MMOL/L, GLUC: 476 MG/DL, CHLORIDE: 112 MMOL/L . E AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 74953 | I-STAT CHEM8+ CARTRIDGE | CHEM8+ CARTRIDGE | JPI | ABBOTT POINT OF CARE | NA | H25301 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |