FDA Adverse Event Malfunction Summary report: N

I-STAT CHEM8+ CARTRIDGE

MDR report key: 24306917 · Received February 10, 2026

Report

Report Number
2245578-2026-00027
Event Type
Malfunction
Date Received
February 10, 2026
Date of Event
January 14, 2026
Report Date
April 13, 2026
Manufacturer
ABBOTT POINT OF CARE
Product Code
JPI
PMA / PMN Number
K183680
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 09-APR-2026. A REVIEW OF THE DEVICE HISTORY RECORDS CONFIRMED THE CARTRIDGE LOTS MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING OF CHEM8+ LOT H25257 AND RETAINED AND RETURNED CARTRIDGE TESTING OF CHEM8+ CARTRIDGE LOT H25301 MET THE ACCEPTANCE CRITERIA PER Q04.01.003 REV. AP, APPENDIX 1 - PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN DETERMINED FOR CHEM8+ CARTRIDGE LOTS H25257 AND H25301.

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4) APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

NA.

Description of Event or Problem · 0

ON (B)(6) 2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CHEM8+ CARTRIDGE THAT YIELDED A DISCREPANT IONIZED CALCIUM RESULT ON AN 8-MONTH-OLD FEMALE PEDIATRIC PATIENT IN THE ICU. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD: DATE: COLLECTED TESTED RESULTS: SAMPLE I-STAT, (B)(6) 2026 , 19:59 , 19:59 , K: 6.8 MMOL/L, GLUC: 166 MG/DL, TCO2: 14 MMOL/L, BUN: 19 MG/DL, ICA: 0.80 MMOL/L A I-STAT, (B)(6) 2026, 20:002 , 20:05 K: 6.9 MMOL/L, GLUC: 210 MG/DL, TCO2: 12 MMOL/L, BUN: 18 MG/DL, ICA: 1.39 MMOL/L B. ROCHE (B)(6) 2026 , 20:02 , 20:30 , K: 7.9 MMOL/L, GLUC: 283 MG/DL, TCO2: 7 MMOL/L, BUN: 14 MG/DL . C. I-STAT , (B)(6) 2026, 21:15 , 21:21 , K: 6.5 MMOL/L, GLUC: 304 MG/DL. CHLORIDE: 120 MMOL/L . D ROCHE (B)(6) 2026 , 21:15, 21:31, K: 9.1 MMOL/L, GLUC: 476 MG/DL, CHLORIDE: 112 MMOL/L . E AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTED THAT PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
74953 I-STAT CHEM8+ CARTRIDGE CHEM8+ CARTRIDGE JPI ABBOTT POINT OF CARE NA H25301

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown