INSET
Report
- Report Number
- 3003442380-2026-00529
- Event Type
- Malfunction
- Date Received
- February 10, 2026
- Date of Event
- January 23, 2026
- Report Date
- February 24, 2026
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244016651
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
CORRECTION: THIS MDR IS BEING SUBMITTED TO CORRECT THE SUBMITTED EXPIRATION DATE UNDER D4 AND MANUFACTURING DATE UNDER H4. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH: A QUERY WAS RUN IN THE EQMS ON 24/FEB/2026 AGAINST "LOT NUMBER" "6011832" AND SIMILAR MALFUNCTION CODES: OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS- BLOCKAGE, OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G., OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED), TUBING IS DAMAGED (E.G., KINKED, DEFORMED, TWISTED, COLLAPSED, OR PINCHED). THE REVIEW CONFIRMED THAT LOT 6011832 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NONCONFORMING REPORTS (NCRS) OR CORRECTIVE AND PREVENTIVE ACTION (CAPAS) OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 24/FEB/2026 AGAINST "LOT NUMBER" CRITERIA EQUAL "6011832" AND SIMILAR MALFUNCTION CODES: OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS- BLOCKAGE, OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G., OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED), TUBING IS DAMAGED (E.G., KINKED, DEFORMED, TWISTED, COLLAPSED, OR PINCHED). THE COUNT OF COMPLAINTS IS 1. THE COMPLAINTS NUMBERS ARE (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011832 WAS MANUFACTURED ACCORDING TO WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN THE LINE INSET 8, ON 19/FEB/2025 WITH A TOTAL OF (B)(4) UNITS. GLUE-TUBING LOT: THE LOT 5B01780 WAS MANUFACTURED ACCORDING TO FORM VERSION 2 AND MANUFACTURED IN THE MACHINE IGTL03, ON 09/FEB/2025 WITH A TOTAL OF (B)(4) UNITS. THE LOT 5B02465 WAS MANUFACTURED ACCORDING TO FORM VERSION 2 AND MANUFACTURED IN THE MACHINE IGTL03, ON 15/FEB/2025 WITH A TOTAL OF (B)(4) UNITS. THE DHR REVIEW INDICATED THAT DURING OUTGOING TEST 4(C), ONE SAMPLE WAS FOUND TO HAVE A LOOSE OBJECT. AN EXTENDED SAMPLING WAS CONDUCTED AND ACCEPTED IN ACCORDANCE WITH ESTABLISHED PROCEDURES. THEREFORE, THE OVERALL REVIEW OF THE DHR CONFIRMS THAT ALL REQUIRED PROCESS-RELATED TESTS WERE COMPLETED AND MET THE APPLICABLE REQUIREMENTS. A NON-CONFORMANCE (NC) WAS FOUND ON REVIEW OF THE DHR WITH NUMBER 2165520, SEVERAL LOADS INTERRUPTED DURING THE STERILIZATION PROCESS BY POWER OUTAGES AT STRINGENCIES G.P. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED IN RELATION TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW IDENTIFIED MINOR FINDINGS. THEY WERE MANAGED ACCORDING TO PROCEDURE AND DID NOT IMPACT COMPLIANCE; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. CAPA DETERMINATION ASSESSMENT - CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI (MONTHLY TRIPS AND ALERTS). CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6011832 AND RELATED MALFUNCTION CODES FOR DETACHMENT. ONE COMPLAINT WAS IDENTIFIED FOR THIS LOT, BUT NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED, AND NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.
REFERENCE NUMBER (B)(4) EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED INSULIN FLOW BLOCKED ALARM DUE TO AN OCCLUSION EVENT ON (B)(6) 2026 WHICH LED TO HYPERGLYCEMIA. THE BLOCKAGE WAS LOCATED IN THE TUBING. THE SITE OF INSERTION WAS ABDOMEN. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 76148 | INSET | UNO INSET I 60/9 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1001681 | 6011832 | 05705244016651 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Female |