FDA Adverse Event Injury Summary report: N

INSET II

MDR report key: 24301217 · Received February 10, 2026

Report

Report Number
3003442380-2026-00365
Event Type
Injury
Date Received
February 10, 2026
Date of Event
January 11, 2026
Report Date
February 10, 2026
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244018129
PMA / PMN Number
K032854
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH: A QUERY WAS RUN IN THE EQMS ON 10-FEB-2026 AGAINST "LOT NUMBER" "6012866" AND SIMILAR MALFUNCTION CODES: SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA BENT (NO HARM), SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - REDUCED FLOW, SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE, SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE. THE REVIEW CONFIRMED THAT LOT 6012866 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NCRS OR CORRECTIVE AND PREVENTIVE ACTION (CAPA)S OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 10-FEB-2026 AGAINST "LOT NUMBER" CRITERIA EQUAL "6012866" AND SIMILAR MALFUNCTION CODES: SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA BENT (NO HARM), SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - REDUCED FLOW, SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE, SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE. THE COUNT OF COMPLAINTS IS 1. THE COMPLAINT NUMBER IS (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6012866 WAS MANUFACTURED ACCORDING TO WORK INSTRUCTION (WI) VERSION 123 AND MANUFACTURED IN THE LINE 7, ON 28-APR-2025 WITH A TOTAL OF (B)(4) UNITS. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED IN ACCORDANCE WITH APPLICABLE PROCEDURES. ALL REQUIRED IN-PROCESS AND FINAL TESTS WERE COMPLETED AND MET SPECIFIED REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED THAT RELATE TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW SUPPORTS COMPLIANCE WITH MANUFACTURING AND QUALITY REQUIREMENTS; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. RETAIN SAMPLES TESTING: RETAIN SAMPLES FROM THE RELEVANT LOT WERE REQUESTED AND TESTED IN ACCORDANCE WITH APPROVED PROCEDURES: WI GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3: ALL 3 SAMPLES TESTED PASSED VISUAL INSPECTION. WI GUIDANCE FOR FUNCTIONAL TESTING 1 AIR FLOW TEST FOR COMPLAINTS AREA VERSION 2: ALL 3 SAMPLES TESTED PASSED. FUNCTIONAL TESTING FOR THE REPORTED MALFUNCTION SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL -BLOCKAGE. ALL TEST RESULTS WERE WITHIN SPECIFICATION AS DOCUMENTED IN THE ATTACHED COMPLAINT (B)(4) TEST REPORT CONCLUSION: TESTING DID NOT CONFIRM THE REPORTED ISSUE; NO NONCONFORMANCE IDENTIFIED. RETURN SAMPLES TESTING: RETURNED SAMPLES FROM THE RELEVANT LOT WERE REQUESTED; HOWEVER, THE CUSTOMER CONFIRMED THAT THE SAMPLE IS NOT AVAILABLE FOR RETURN. CONCLUSION: VISUAL AND FUNCTIONAL TESTING COULD NOT BE PERFORMED DUE TO LACK OF SAMPLE AVAILABILITY. ASSESSMENT WILL BE BASED ON OTHER AVAILABLE EVIDENCE. CAPA DETERMINATION ASSESSMENT - CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI (MONTHLY TRIPS AND ALERTS). COMPLAINT UNCONFIRMED: FOLLOWING INVESTIGATION THE REPORTED FAILURE COULD NOT BE CONFIRMED FOR THIS COMPLAINT. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING. A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6012866 AND RELATED MALFUNCTION CODES. ONE COMPLAINT WAS IDENTIFIED FOR THIS LOT; HOWEVER, NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. SAMPLES WERE REQUESTED; HOWEVER, THE CUSTOMER CONFIRMED THAT NO SAMPLES WERE AVAILABLE FOR ANALYSIS. CONSEQUENTLY, AN INVESTIGATION WAS CONDUCTED USING REFERENCE SAMPLES, AND NO FAILURES RELATED TO THE COMPLAINT WERE IDENTIFIED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION AND REFERENCE SAMPLE ANALYSIS ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4) EVENT OCCURRED IN SPAIN. IT WAS REPORTED THAT PATIENT FACED INFUSION SET BENT CANNULA DUE TO WHICH THE PATIENT FACED HYPERGLYCEMIA EVENT ON (B)(6) 2026.THE PATIENT WENT TO EMERGENCY ROOM AND THE DURATION OF EMERGENCY ROOM STAY WAS FOR 11 HOURS. THE BLOOD GLUCOSE LEVEL WAS 411 MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS AND FLUIDS OF SALINE AND INSULIN. THE PATIENT WAS TESTED POSITIVE FOR KETONES AND HAD MODERATE KETONES. THE PATIENT WAS RELEASED FROM THE HOSPITAL ON (B)(6) 2026. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
363059 INSET II UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002817 6012866 05705244018129

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention