COBAS INTEGRA 400 PLUS
Report
- Report Number
- 1823260-2012-00543
- Event Type
- Malfunction
- Date Received
- January 26, 2012
- Date of Event
- January 13, 2012
- Report Date
- January 26, 2012
- Manufacturer
- ROCHE DIAGNOSTICS
- Product Code
- JGS
- PMA / PMN Number
- K951595
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE USER DISCOVERED THE ACTIVATOR WAS EMPTY WHICH SHE BELIEVED WAS THE CAUSE OF THE ISSUE.
THE USER NOTICED LOW ANION GAPS FOR PATIENT SAMPLES AND FOUND QUESTIONABLE ION SELECTIVE ELECTRODE (ISE) RESULTS FOR TWO PATIENT SAMPLES. OF THE DATA PROVIDED, THE SODIUM AND CHLORIDE RESULTS FOR ONE PATIENT SAMPLES WERE DISCREPANT AND REPORTED OUTSIDE THE LABORATORY. THE INITIAL SODIUM RESULT WAS 133 MMOL/L AND THE INITIAL CHLORIDE RESULT WAS 101MMOL/L WHICH WERE REPORTED OUTSIDE THE LABORATORY. THE SAMPLE WAS REPEATED AND THE SODIUM RESULT WAS 138 MMOL/L AND THE CHLORIDE RESULT WAS 104 MMOL/L. THE USER THEN RAN PRECISION TESTING ON THE PATIENT SAMPLE. THE SODIUM RESULTS WERE 137, 134, 139, 134, 129, 130, 133, 130 AND 129 MMOL/L. THE CHLORIDE RESULTS WERE 102, 99, 103, 99, 96, 97, 99, 98 AND 96 MMOL/L. THE PATIENT WAS NOT TREATED BASED ON THE ERRONEOUS RESULTS REPORTED AND THE USER SENT A CORRECTED REPORT WITH THE REPEAT RESULTS. THE SODIUM ELECTRODE LOT NUMBER WAS 21511250 WITH AN EXPIRATION DATE OF 12/30/2011. THE CHLORIDE ELECTRODE LOT NUMBER AND EXPIRATION DATE WAS NOT PROVIDED. THE USER DISCOVERED THE ACTIVATOR WAS EMPTY WHICH SHE BELIEVED WAS THE CAUSE OF THE ISSUE. SHE REPLACED THE ACTIVATOR AND REPEATED THE PRECISION TESTING USING A DIFFERENT SAMPLE AND ALL SODIUM RESULTS AGREED WITHIN 2 MMOL/L.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COBAS INTEGRA 400 PLUS | CLINICAL CHEMISTRY ANALYZER | JGS | ROCHE DIAGNOSTICS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |