FDA Adverse Event Injury Summary report: N

KIT IMPLANTABLE SLIM TIP LEAD, 50CM

MDR report key: 24293536 · Received February 9, 2026

Report

Report Number
1627487-2026-00633
Event Type
Injury
Date Received
February 9, 2026
Date of Event
January 21, 2026
Report Date
May 29, 2026
Manufacturer
ABBOTT MEDICAL
Product Code
PMP
UDI-DI
05415067027153
PMA / PMN Number
P150004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE ALLEGATION IS AGAINST 1 OF 2 LEADS; HOWEVER, IT IS UNKNOWN WHICH LEAD(S), THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: LEAD, MODEL: MN10450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 8060953.

Additional Manufacturer Narrative · 0

CORRECTION TO: THE ALLEGATION IS AGAINST 1 OF 2 LEADS; HOWEVER, IT IS UNKNOWN WHICH LEAD(S), THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: LEAD, MODEL: MN10450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 8060953.

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A LEAD REVISION ON (B)(6) 2026 [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-00630], [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-00631], [RELATED MANUFACTURER REFERENCE NUMBER: 1627487-2026-00632], THE LEFT L1 LEAD BROKE LEAVING A FRAGMENT IN THE PATIENT. SURGICAL INTERVENTION MAY TAKE PLACE IN THE FUTURE TO ADDRESS THE ISSUE. IT IS NOT KNOWN WHICH L1 LEAD REMAINS PARTIALLY IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156999 KIT IMPLANTABLE SLIM TIP LEAD, 50CM DRG LEAD PMP ABBOTT MEDICAL MN10450-50A 8060953 05415067027153

Patients

Seq Age Sex Outcome Treatment
1