FDA Adverse Event Malfunction Summary report: N

PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD

MDR report key: 24290826 · Received February 9, 2026

Report

Report Number
2183870-2026-00060
Event Type
Malfunction
Date Received
February 9, 2026
Date of Event
January 13, 2026
Report Date
April 23, 2026
Manufacturer
COVIDIEN
Product Code
FGE
UDI-DI
00763000066390
PMA / PMN Number
P110023
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

PRODUCT ANALYSIS THE DEVICE WAS RETURNED WITH THE LOCK PIN MECHANISM TIGHTENED AND WITH THE STENT ENCLOSED, THE DEVICE WAS IDENTIFIED FROM THE STRAIN RELIEF AS 60MM WITH 5X60MM MENTIONED, THE DEVICE WAS RETURNED WITH THE POUCH LABEL WITH MODEL NUMBER MENTIONED AS PRP35-06-060-120 AND DIAMETER AS 6MM, THE STENT WAS DEPLOYED BY ADVANCING THE PUSH GRIPS AND THE DEPLOYED STENT WAS MEASURED AND CONFIRMED AS 60MM WITH NO ABNORMALITIES OBSERVED, THE DEPLOYED STENT DIAMETER WAS MEASURED WITH RULER AND CONFIRMED AS 5MM, THE DEPLOYED STENT DIAMETER WAS ALSO MEASURED USING LASER MICROMETRE AS 0.211 INCHES WHICH IS APPROXIMATELY 5.3MM, MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED THAT AN ATTEMPT WAS MADE TO USE A PROTEGE EF TO TREAT A CALCIFIED LESION IN THE MID SUPERFICIAL FEMORAL ARTERY. DEGREE OF TORTUOSITY WAS MODERATE. DEGREE OF CALCIFICATION WAS MODERATE. LESION WAS PRE-DILATED WITH A  5-60 DEVICE. AFTER OPENING THE PACKAGING, IT WAS FOUND THAT THE STENT MODEL INDICATED ON THE STENT DID NOT MATCH THE MODEL ON THE OUTER PACKAGING. A NEW PRP35-06-060-120 STENT WAS USED TO COMPLETE PROCEDURE. NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
350787 PROTEGE EVERFLEX SELF-EXPANDING STENT SYSTEM STANDARD CATHETER, BILIARY, DIAGNOSTIC FGE COVIDIEN C161965 00763000066390

Patients

Seq Age Sex Outcome Treatment
1 56 YR Male