FDA Adverse Event Other Summary report: N

ABL80 FLEX CO-OX ANALYZER

MDR report key: 2428876 · Received January 21, 2012

Report

Report Number
2027541-2012-00003
Event Type
Other
Date Received
January 21, 2012
Date of Event
October 27, 2011
Report Date
November 15, 2011
Manufacturer
SENDX MEDICAL, INC.
Product Code
CHL
PMA / PMN Number
K080370
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DATA LOGS WERE OBTAINED FROM THE ANALYZER, AND HAVE BEEN INVESTIGATED. THE CALIBRATION AND QUALITY CONTROL RECORDS FOR THIS ANALYZER DEMONSTRATE THE ANALYZER TO BE PERFORMING PROPERLY. THE PT SAMPLE WITH A GLUCOSE MEASURED VALUE OF 5.2 MMOL/L ALSO HAD AN OXYGEN VALUE OF 15 MMHG. THE ABL80 REFERENCE MANUAL PROVIDES PERFORMANCE SPECIFICATIONS FOR THE MEASUREMENT OF GLUCOSE. THE MANUAL DEFINES THE LIMITS OF GLUCOSE LINEARITY AS A FUNCTION OF THE AVAILABLE OXYGEN IN THE SAMPLE. THERE ARE NO LINEARITY CLAIMS WHEN THE OXYGEN LEVEL IS BELOW 20 MMHG.

Description of Event or Problem · 1

THE USER REPORTS THAT THE ABL80 RETURNED A GLUCOSE RESULT OF 5.2 MMOL/L. THE SAMPLE WAS RETESTED ON THE ABL80 AND THE GLUCOSE RESULT WAS 5.4 MMOL/L. THE ABL80 GLUCOSE RESULTS DID NOT FIT THE CLINICAL STATUS OF THE DIABETIC KETOACIDOSIS PT AND GLUCOSE RESULT FROM A GLUCOMETER. THE USER THEN SENT THE SAMPLE TO THE LABORATORY WHICH REGISTERED A GLUCOSE RESULT OF 54.8 MMOL/L. THERE IS NO PT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ABL80 FLEX CO-OX ANALYZER PH,BLOODGAS,ELECTROLYTE ANALYSIS SYSTEM CHL SENDX MEDICAL, INC. 393-841

Patients

Seq Age Sex Outcome Treatment
1