FDA Adverse Event Malfunction Summary report: N

PRESTIGE CERVICAL DISC SYSTEM

MDR report key: 24288729 · Received February 9, 2026

Report

Report Number
1824199-2026-00015
Event Type
Malfunction
Date Received
February 9, 2026
Date of Event
January 16, 2026
Report Date
March 18, 2026
Manufacturer
WARSAW ORTHOPEDIC, INC.
Product Code
MJO
UDI-DI
00613994714855
PMA / PMN Number
P090029
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

D4: LOT NUMBER IS UNKNOWN MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS OF PRODUCT:6973060, LOT:EM17G011 VISUAL AND OPTICAL INSPECTION CONFIRMED THE INSERTER IS WORN FROM REPEATED USE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF) PROCEDURE AT LEVELS 5-7 FOR A PRE-OPERATIVE DIAGNOSIS OF HERNIATED NUCLEUS PULPOSUS. IT WAS REPORTED THAT THREE INSERTERS HAVE BEEN BROKEN AND NO LONGER ATTACHED TO THE IMPLANTS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE MANUFACTURER REPRESENTATIVE THAT THERE WAS AN INSTRUMENT INCOMPATIBILITY WHICH MIGHT BE DUE TO WEAR ISSUE AND NOT BREAKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
89928 PRESTIGE CERVICAL DISC SYSTEM PROSTHESIS, INTERVERTEBRAL DISC MJO WARSAW ORTHOPEDIC, INC. 6973070 EM17G011 00613994714855

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown