FDA Adverse Event Malfunction Summary report: N

DAVINCI XI

MDR report key: 24286154 · Received February 8, 2026

Report

Report Number
2955842-2026-03685
Event Type
Malfunction
Date Received
February 8, 2026
Date of Event
January 13, 2026
Report Date
March 8, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874110744
PMA / PMN Number
K131861
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ANNEX CODES C AND D HAVE BEEN CHANGED.

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO FURTHER INVESTIGATE THE REPORTED COMPLAINT. THE FSE WAS NOT ABLE TO REPRODUCE THE ISSUE BUT DID MAKE ELECTRICAL MECHANICAL ADJUSTMENTS. NO PARTS WERE REPLACED. SYSTEM WAS TESTED AND VERIFIED AS READY FOR USE.

Additional Manufacturer Narrative · 0

AN INTUITIVE SURGICAL, INC. (ISI) FIELD SERVICE ENGINEER (FSE) HAS BEEN REQUESTED TO INVESTIGATE THE REPORTED EVENT. AT THIS TIME, THE FSE INVESTIGATION HAS NOT BEEN COMPLETED.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED LOW ANTERIOR RESECTION SURGICAL PROCEDURE WITH DUAL CONSOLE, THE CUSTOMER INFORMED AN INTUITIVE SURGICAL, INC. (ISI) TECHNICAL SUPPORT ENGINEER (TSE) THAT A NON-RECOVERABLE ERROR 40067 APPEARED. THE TSE CONFIRMED ERROR 40067 IN THE LOGS POINTING TO THE SECOND SURGEON SIDE CONSOLE (SSC). THE CUSTOMER CONTINUED AND COMPLETED THE PROCEDURE WITH THE OTHER SSC. THERE WAS NO REPORTED INJURY. THE PROCEDURE WAS NOT DELAYED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
348640 DAVINCI XI SURGEON SIDE CONSOLE, SMART PEDALS NAY INTUITIVE SURGICAL, INC 380677-28 N/A 00886874110744

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES