FDA Adverse Event
Injury
Summary report: N
R-HEAD RECON HEAD
MDR report key: 2428566
·
Received January 20, 2012
Report
- Report Number
- 3003640913-2012-00006
- Event Type
- Injury
- Date Received
- January 20, 2012
- Date of Event
- December 23, 2011
- Report Date
- January 20, 2012
- Manufacturer
- SMALL BONE INNOVATIONS, INC.
- Product Code
- KWI
- PMA / PMN Number
- K023604
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
REVIEW OF SURGERY REPORT STATES THAT HEAD HAD BEEN WORN FOR 4 YEARS AND THAT THE HEAD DISPLACES THEN THE PT FELL. EVALUATION OF EXPLANTED HEAD SHOWED WEAR AND MARKS FROM BEING DISASSOCIATED FROM STEM. ALSO SHOWED MARKS OF IMPACTION POSSIBLY FROM FALL.
Description of Event or Problem · 1
R-HEAD RECON HEAD WAS REVISED IN A PT AFTER A FALL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | R-HEAD RECON HEAD | RADIAL HEAD PROSTHESIS | KWI | SMALL BONE INNOVATIONS, INC. | RCN-H3 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 40 YR | Other |