Q-SYTE
Report
- Report Number
- 1710034-2026-00116
- Event Type
- Malfunction
- Date Received
- February 6, 2026
- Date of Event
- January 28, 2026
- Report Date
- February 10, 2026
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FPA
- UDI-DI
- 00382903851003
- PMA / PMN Number
- K013621
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED.
A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 385100 AND LOT NUMBER 5062596. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED ISSUE, AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS A SAMPLE WAS UNAVAILABLE FOR RETURN, A THOROUGH SAMPLE INVESTIGATION COULD NOT BE COMPLETED. BASED ON THE INVESTIGATION RESULTS, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED.THERE ARE QUALITY CONTROLS CURRENTLY IN PLACE TO DETECT SIMILAR ISSUES DURING THE PRODUCTION PROCESS. FURTHER ACTION HAS NOT BEEN DETERMINED NECESSARY AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
AFTER CONNECTING THE INFUSION SET TO THE PATIENT, LEAKAGE WAS DETECTED. UPON DISCONNECTING THE SET, IT WAS OBSERVED THAT THE RUBBER CONTACT SURFACE AT THE CONNECTOR HAD BECOME INDENTED.
NO NEW INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 345360 | Q-SYTE | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 5062596 | 00382903851003 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |