FDA Adverse Event Malfunction Summary report: N

PRN ADAPTER

MDR report key: 24279227 · Received February 6, 2026

Report

Report Number
3006948883-2026-00067
Event Type
Malfunction
Date Received
February 6, 2026
Date of Event
January 4, 2026
Report Date
March 4, 2026
Manufacturer
BD SUZHOU (MDS)
Product Code
FMG
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: A DEVICE HISTORY REVIEW WAS CONDUCTED FOR LOT NUMBER 5203385. OUR RECORDS SHOW THAT THIS IS THE ONLY INSTANCE OF THIS ISSUE OCCURRING IN THIS PRODUCTION BATCH. ACCORDING TO THE SAMPLING PLAN APPLIED FOR PRODUCT PERFORMANCE, THIS LOT WAS ACCEPTED AND RELEASED WITHOUT DEFECTS BEING NOTED DURING THE FINAL ASSEMBLY OR VISUAL INSPECTIONS. A SAMPLE COULD NOT BE OBTAINED FOR EVALUATION AND TESTING; IN LIEU OF THE AFFECTED DEVICE, FUNCTIONAL TESTING WAS PERFORMED ON RETENTION SAMPLES FOR THIS LOT, THE RESULTS OF THESE SHOW THAT THE TESTED UNITS PERFORMED WITHIN PRODUCT SPECIFICATIONS. UNFORTUNATELY, WITHOUT THE ABILITY TO INVESTIGATE THE AFFECTED UNIT OUR QUALITY ENGINEERS WERE UNABLE TO DETERMINE THE ROOT CAUSE FOR THIS COMPLAINT.

Description of Event or Problem · 0

ON (B)(6) 2026, 1:40 PM: WHILE CHANGING BED LINENS, THE PATIENT'S FAMILY NOTICED DAMP CLOTHING AND IMMEDIATELY CALLED THE ON-DUTY NURSE. THE NURSE PROMPTLY ARRIVED TO INVESTIGATE AND CONFIRMED THE DAMPNESS RESULTED FROM A SLOW LEAK OF MEDICATION (FAT EMULSION AMINO ACID GLUCOSE INJECTION) AT THE HEPARIN CAP CONNECTED TO THE PATIENT'S IV PORT NEEDLE. THIS HEPARIN CAP HAD BEEN NEWLY REPLACED AT 8:00 AM THAT SAME DAY. THE NURSE PROMPTLY REPLACED THE HEPARIN CAP WITH A NEW STERILE ONE, ENSURING A SECURE CONNECTION AND UNOBSTRUCTED TUBING. THIS INCIDENT RESULTED IN MEDICATION WASTE BUT CAUSED NO HARM TO THE PATIENT AT THIS TIME.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
337144 PRN ADAPTER STOPCOCK, I.V. SET FMG BD SUZHOU (MDS) 5203385

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown