FDA Adverse Event Other Summary report: N

SONIC GENERIC ULTRASOUND TRANSMISSION GEL

MDR report key: 2427803 · Received January 19, 2012

Report

Report Number
2241583-2012-00001
Event Type
Other
Date Received
January 19, 2012
Date of Event
July 21, 2006
Report Date
July 21, 2006
Manufacturer
PHARMACEUTICAL INNOVATIONS, INC.
Product Code
GYB
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

PHARMACEUTICAL INNOVATIONS RECEIVED A COMPLAINT ON (B)(4) 2006, FROM (B)(6) MEDICAL CENTER, (B)(4). ON (B)(4) 2006, AN MDR WAS RECEIVED AT PHARMACEUTICAL INNOVATIONS FROM THE FDA. PHARMACEUTICAL INNOVATIONS HEAD CHEMIST SPOKE WITH (B)(6) AT (B)(6) MEDICAL CENTER ON (B)(4) 2006, AND SHE CONFIRMED THE CONTAMINATION WAS NOT FROM PHARMACEUTICAL INNOVATION'S GEL. THE PART NUMBER AND LOT NUMBER REFERENCED ON THE MDR WAS NOT PHARMACEUTICAL INNOVATIONS, INC. NOMENCLATURE FOR PART NUMBER OR LOT NUMBER. THE PART NUMBER P11111-05 IS MANUFACTURED BY JOH ALI MEDICAL, INC. THIS MDR WAS NEVER COMPLETED AND SUBMITTED TO THE FDA. THE HEAD CHEMIST DID NOT SUBMIT THE NECESSARY DOCUMENTATION IN 2006. AS A RESULT OF A RECENT FDA INSPECTION THIS ISSUE WAS BROUGHT TO THE ATTENTION OF PHARMACEUTICAL INNOVATIONS, INC. MANAGEMENT AND IS NOW BEING ADDRESSED.

Description of Event or Problem · 1

EVENT DESC: INFECTION CONTROL NOTED THAT NICU BABIES WITH POSITIVE CULTURES FOR PSEUDOMONAS HAD ALL GONE FOR ECHOCARDIOGRAMS. INVESTIGATION LEAD TO CULTURING THE GEL. ULTRASOUND GEL COMES IN 5 LITERS CONTAINERS WITH 8 OZ REFILLABLE BOTTLES. BOTTLES ARE TOP OFF AND NOT CAPPED. EIGHT OUNCES BOTTLES AND 5 LITER BOTTLES TESTED POSITIVE FOR THE PSEUDOMONAS ORGANISM. A NEW UNOPENED CONTAINER WAS SENT OUT FOR CULTURE - RESULTS ARE PENDING. FOUR AREAS OF THE HOSPITAL USE THE REFILLABLE BOTTLES AND REST OF THE HOSPITAL BUYS THE 8 OZ BOTTLES. INFECTION CONTROL DISCOURAGES TOPPING OFF ANY FLUIDS. IT IS NOT KNOWN FOR SURE IF THE ULTRASOUND GEL WAS THE SOURCE OF THE NICU BABIES INFECTIONS AS IS A COMMON ORGANISM. INVESTIGATION IS ONGOING. DEVICE USAGE PROBLEM: OTHER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SONIC GENERIC ULTRASOUND TRANSMISSION GEL GEL, ULTRASOUND GYB PHARMACEUTICAL INNOVATIONS, INC. UNK 122163

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization| R