DAVINCI XI
Report
- Report Number
- 2955842-2026-03383
- Event Type
- Malfunction
- Date Received
- February 6, 2026
- Date of Event
- January 16, 2026
- Report Date
- March 25, 2026
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- UDI-DI
- 00886874110720
- PMA / PMN Number
- K131861
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INTUITIVE SURGICAL INC. (ISI) RECEIVED THE UNIVERSAL SURGICAL MANIPULATOR (USM) 4 FOR FAILURE ANALYSIS INVESTIGATION. THE UNIT WAS ANALYZED AND IN THE SYSTEM LOGS, NO DATA WAS FOUND TO INDICATE THAT THE FAULT HAD OCCURRED IN THE FIELD. UPON VISUAL INSPECTION, NO ISSUES WERE FOUND THAT WOULD BE RELATED TO THE REPORTED EVENT. THE USM WAS INSTALLED ONTO A GOLDEN SYSTEM WHERE THE COMPONENT FUNCTIONED AS EXPECTED. THE USM WAS THEN INSTALLED ONTO A PATIENT SIDE CART FIXTURE TEST PLATFORM (PFTP) WHERE ALL RELEVANT TESTING WAS PASSED WITHIN SPECIFICATION. ONCE TESTING WAS COMPLETED, ALL SEARCHLIGHT, CHIP-ENCODER VIRTUAL ABSOLUTE (CVA), AND HALL SENSOR ASSEMBLIES WERE INSPECTED, BUT NO FAULTS COULD BE IDENTIFIED. THE COMPLAINT WAS NOT CONFIRMED BASED ON FAILURE ANALYSIS.
INTUITIVE SURGICAL INC. (ISI) HAS NOT RECEIVED THE DA VINCI PRODUCT WITH AN ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE UNIVERSAL SURGICAL MANIPULATOR (USM) 4 WAS NOT WORKING. THE TECHNICAL SERVICE ENGINEER WAS UNABLE TO VIEW THE SYSTEM LOGS DURING THE CALL AND NO FURTHER INFORMATION WAS AVAILABLE. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 255669 | DAVINCI XI | PATIENT SIDE CART, 4-ARM | NAY | INTUITIVE SURGICAL, INC | 380652-51 | N/A | 00886874110720 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES. |