QUICK SET
Report
- Report Number
- 3003442380-2026-00359
- Event Type
- Malfunction
- Date Received
- February 6, 2026
- Date of Event
- January 16, 2026
- Report Date
- March 18, 2026
- Manufacturer
- UNOMEDICAL UM-D
- Product Code
- FPA
- UDI-DI
- 05705244017573
- PMA / PMN Number
- K160648
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR (B)(4) - DEVICE 1 OF 2.
SUPPLEMENTAL REPORT 01 (B)(4) MDR 3003442380-2026-00359. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS, H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH: A QUERY WAS RUN IN THE EQMS ON 17/MAR/2026 AGAINST "LOT NUMBER" "6002729" AND SIMILAR MALFUNCTION CODES: OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS- BLOCKAGE, OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G., OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED), TUBING IS DAMAGED (E.G., KINKED, DEFORMED, TWISTED, COLLAPSED, OR PINCHED). THE REVIEW CONFIRMED THAT LOT 6002729 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NONCONFORMING REPORTS (NCR)S OR CORRECTIVE AND PREVENTIVE ACTION (CAPA)S OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 17/MAR/2026 AGAINST "LOT NUMBER" CRITERIA EQUAL "6002729" AND SIMILAR MALFUNCTION CODES: OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS- BLOCKAGE, OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G., OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED), TUBING IS DAMAGED (E.G., KINKED, DEFORMED, TWISTED, COLLAPSED, OR PINCHED). THE COUNT OF COMPLAINTS IS 2. THE COMPLAINTS NUMBERS ARE (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6002729 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 77 AND MANUFACTURED IN THE MACHINE MULTIVAC 12, ON 14/AUG/2023 WITH A TOTAL OF (B)(4) UNITS. GLUE-TUBING LOT: THE LOT 3H00361 WAS MANUFACTURED ACCORDING TO WI VERSION 38 AND MANUFACTURED IN THE MACHINE MP08, ON 13/AUG/2023 WITH A TOTAL OF (B)(4) UNITS. THE LOT 3H00364 WAS MANUFACTURED ACCORDING TO WI VERSION 38 AND MANUFACTURED IN THE MACHINE MP04, ON 13/AUG/2023 WITH A TOTAL OF (B)(4) UNITS. THE LOT 3H00363 WAS MANUFACTURED ACCORDING TO WI VERSION 38 AND MANUFACTURED IN THE MACHINE MP08, ON 12/AUG/2023 WITH A TOTAL OF (B)(4) UNITS. THE LOT 3H00365 WAS MANUFACTURED ACCORDING TO WI VERSION 38 AND MANUFACTURED IN THE MACHINE MP05, ON 11/AUG/2023 WITH A TOTAL OF (B)(4) UNITS. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED IN ACCORDANCE WITH APPLICABLE PROCEDURES. ALL REQUIRED IN-PROCESS AND FINAL TESTS WERE COMPLETED AND MET SPECIFIED REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED THAT RELATE TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW SUPPORTS COMPLIANCE WITH MANUFACTURING AND QUALITY REQUIREMENTS; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. CAPA DETERMINATION ASSESSMENT - CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI (MONTHLY TRIPS AND ALERTS). COMPLAINT UNCONFIRMED: FOLLOWING INVESTIGATION THE REPORTED FAILURE COULD NOT BE CONFIRMED FOR THIS COMPLAINT. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6002729 AND RELATED MALFUNCTION CODES FOR OCCLUSION. TWO COMPLAINTS WERE IDENTIFIED FOR THIS LOT, BUT NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED, AND NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. THIS EMDR IS BEING SUBMITTED FOR UNKNOWN NUMBER OF QUANTITIES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2026. THE BLOCKAGE WAS IN THE TUBING. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 22486 | QUICK SET | UNO QUICK-SET 60/9 SC1 MECA | FPA | UNOMEDICAL UM-D | MMT-397A | 6002729 | 05705244017573 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |