RUGGLES LEYLA RETRACTION SYSTEM - BALL JOINT CLA
Report
- Report Number
- 2523190-2012-00005
- Event Type
- Injury
- Date Received
- January 9, 2012
- Date of Event
- December 21, 2011
- Report Date
- January 9, 2012
- Manufacturer
- INTEGRA YORK, PA, INC.
- Product Code
- GZT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
TO DATE THE DEVICE INVOLVED IN THE REPORTED INCIDENT HAS NOT BEEN RECEIVED FOR EVALUATION. AN INVESTIGATION HAS BEEN INITIATED BASED ON THE REPORTED INFORMATION.
"CLAMP BALL JOINT FAILED WHEN HOOKS WERE PLACED CAUSING PATIENT INJURY." ADDITIONAL INFORMATION RECEIVED FROM THE ASSISTANT NURSE MANAGER NOTED THAT DURING A RIGHT SIDED CRANIOTOMY PERFORMED FOR THE RESECTION OF AN ARTERIAL VENOUS MALFORMATION PERFORMED ON (B)(6) 2011, THE POSITIONING ARM OF THE LEYLA RETRACTION SYSTEM SLIPPED CAUSING A LACERATION TO THE MIDDLE CEREBRAL ARTERY. THE TREATMENT WAS A REPAIR OF THE LACERATION TO THE MIDDLE CEREBRAL ARTERY. THE SURGERY TIME WAS PROLONGED BY 10-15 MINUTES FOR THE REPAIR. THE SURGICAL OUTCOME WAS REPORTED AS "GOOD." PER THE DIRECTOR OF BIOMEDICAL ENGINEERING A MEDWATCH REPORT #(B)(4) WAS SUBMITTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RUGGLES LEYLA RETRACTION SYSTEM - BALL JOINT CLA | NA | GZT | INTEGRA YORK, PA, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 15 YR | Life Threatening| R | HEAD| HOLDING ROD| R2381 RUGGLES LEYLA RETRACTION SYSTEM - RIGID| R2382 RUGGLES LEYLA RETRACTION SYSTEM - COUPLING |