- Report Number
- 6000001-2012-04353
- Event Type
- Malfunction
- Date Received
- January 25, 2012
- Date of Event
- November 30, 2011
- Report Date
- December 27, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 589:317:1061:0000 WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO DEPLETED MAIN BATTERIES. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER PERSONNEL, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
(B)(4). UPON FURTHER INVESTIGATION BY QUALITY ENGINEERING, IT WAS DETERMINED THAT THERE IS NO INFORMATION TO SUGGEST THAT THIS DEVICE MALFUNCTION WOULD CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF IT WERE TO RECUR.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE OF 589:317:1061:0000. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORTED PATIENT INVOLVEMENT. THERE WAS NO REPORT OF PATIENT/USER INJURY OR MEDICAL INTERVENTION NEEDED IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |