FDA Adverse Event Malfunction Summary report: N

MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL

MDR report key: 24268153 · Received February 5, 2026

Report

Report Number
2032227-2026-121234
Event Type
Malfunction
Date Received
February 5, 2026
Date of Event
January 15, 2026
Report Date
April 9, 2026
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
PMA / PMN Number
P160017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

UNIT PASSED DISPLACEMENT TEST, REWIND TEST, PRIME/SEATING TEST, AND BASIC OCCLUSION TEST. FAILED THE FORCE SENSOR TEST DUE TO UNEXPECTED INSULIN FLOW BLOCKED ALARM CAUSED BY MOISTURE DAMAGE TO THE FORCE SENSOR. UNABLE TO PERFORM OCCLUSION TEST DUE TO UNEXPECTED INSULIN FLOW BLOCKED ALARM. UNIT SUCCESSFULLY DOWNLOADED TO THUMP. CONFIRMED PUMP ALARMED INSULIN FLOW BLOCKED ALARM DURING NORMAL BOLUS ON 01/14/2026 17:37:43 TO 01/14/2026 17:36:34 IN PUMP DOWNLOADED HISTORY. FOUND MOISTURE DAMAGE TO MOTOR ASSEMBLY, PCB1 BOARD AND PCB 2 BOARD DURING VISUAL INSPECTION. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: CORRODED ELECTRONIC ASSEMBLIES, CORRODED BATTERY TUBE, CORRODED MOTOR HOME SWITCH, SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, CRACKED CASE BATTERY TUBE, CRACKED BATTERY TUBE THREADS, AND CRACKED KEYPAD OVERLAY. THE P-CAP/RESERVOIR DOES LOCK PROPERLY. CONFIRMED UNEXPECTED INSULIN FLOW BLOCKED ALARM DURING ANALYSIS DUE TO MOISTURE DAMAGE TO THE FORCE SENSOR. CONFIRMED MOISTURE DAMAGE TO MOTOR ASSEMBLY, PCB1 BOARD AND PCB 2 BOARD DURING VISUAL INSPECTION. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA, HYPOGLYCEMIA AND CUSTOMER ALLEGED INSULIN FLOW BLOCK. THE EVENT INVOLVED PRODUCT MMT-1884. TROUBLESHOOTING WAS PERFORMED AND IT WAS UNKNOWN WHETHER THE CUSTOMER HAD BEEN USING THE INSULIN PUMP WITHIN 48 HOURS OF THE REPORTED EVENT, AND IT WAS UNKNOWN WHETHER THE AUTO MODE FEATURE WAS ACTIVE AT THE TIME OF THE EVENT. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED. THE CUSTOMER WILL DISCONTINUE PUMP USE AND REVERT TO BACK UP PLAN AS PER HEALTH CARE PROFESSIONAL INSTRUCTIONS. THE PRODUCT MMT-1884 WILL BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330251 MINIMED 780G US SYSTEM BLE CONNECT 3.0 MG/DL AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1884 NG3681053H

Patients

Seq Age Sex Outcome Treatment
1 73 YR Female