BD VACUTAINER® SST¿
Report
- Report Number
- 1024879-2026-00107
- Event Type
- Malfunction
- Date Received
- February 4, 2026
- Date of Event
- January 9, 2026
- Report Date
- February 24, 2026
- Manufacturer
- BECTON DICKINSON & CO (FRANKLIN LAKES)
- Product Code
- JKA
- UDI-DI
- 30382903679837
- PMA / PMN Number
- BK050036
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: E.1. INITIAL REPORTER PHONE # (B)(6). H.4. DEVICE MANUFACTURE DATE: 01-JUN-2025. INVESTIGATION SUMMARY: BD RECEIVED 3 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND UNDERFILL WAS OBSERVED IN LOT 5171519. ADDITIONALLY, 20 RETAINED SAMPLES FROM EACH LOT WERE FUNCTIONALLY TESTED FOR LOW OR NO DRAW AND ALL 60 TUBES WERE FOUND TO BE WITHIN SPECIFICATION. THIS COMPLAINT HAS BEEN CONFIRMED FOR LOT 5171519, WITH RESPECT TO UNDERFILL. BD WAS UNABLE TO IDENTIFY A DEFINITIVE ROOT CAUSE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
E.1. INITIAL REPORTER FACILITY NAME: (B)(6) HOSPITAL. THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K230855. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
REPORT 1 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® SST¿, TWO TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
REPORT 1 OF 3, IT WAS REPORTED WHILE USING BD VACUTAINER® SST¿, TWO TUBES UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 4204 | BD VACUTAINER® SST¿ | TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION | JKA | BECTON DICKINSON & CO (FRANKLIN LAKES) | 5171519 | 30382903679837 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |