FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿

MDR report key: 24262700 · Received February 4, 2026

Report

Report Number
1024879-2026-00108
Event Type
Malfunction
Date Received
February 4, 2026
Date of Event
January 9, 2026
Report Date
February 24, 2026
Manufacturer
BECTON DICKINSON & CO (FRANKLIN LAKES)
Product Code
JKA
UDI-DI
30382903679837
PMA / PMN Number
BK050036
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E.1. INITIAL REPORTER PHONE # (B)(6). INVESTIGATION SUMMARY: BD RECEIVED 3 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND UNDERFILL WAS NOT OBSERVED WITH LOT 5140352. ADDITIONALLY, 20 RETAINED SAMPLES FROM EACH LOT WERE FUNCTIONALLY TESTED FOR LOW OR NO DRAW AND ALL 60 TUBES WERE FOUND TO BE WITHIN SPECIFICATION. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR LOT 5140352 WITH RESPECT TO UNDERFILL. BD WAS UNABLE TO IDENTIFY A DEFINITIVE ROOT CAUSE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

E.1. INITIAL REPORTER FACILITY NAME: (B)(6). THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K230855. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

REPORT 2 OF 3: IT WAS REPORTED WHILE USING BD VACUTAINER® SST¿, ONE TUBE UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

REPORT 2 OF 3. IT WAS REPORTED WHILE USING BD VACUTAINER® SST¿, ONE TUBE UNDERFILLED. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
4198 BD VACUTAINER® SST¿ TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION JKA BECTON DICKINSON & CO (FRANKLIN LAKES) 5140352 30382903679837

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown