FDA Adverse Event
Other
Summary report: N
STARBURST TALON SEMI-FLEX
MDR report key: 2426260
·
Received January 17, 2012
Report
- Report Number
- 1056436-2012-00006
- Event Type
- Other
- Date Received
- January 17, 2012
- Date of Event
- November 17, 2011
- Report Date
- January 17, 2011
- Manufacturer
- ANGIODYNAMICS
- Product Code
- GEI
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
DEVICE APPEARED TO FUNCTION PROPERLY AT TIME OF SURGERY ((B)(6) 2011) 2.9 CM TUMOR INFERIOR LEFT KIDNEY RF ABLATION WITH 3CM AND 4CM DEPLOYMENT TINE CLOSEST TO URETER (2CM) AT 3CM DEPLOYMENT (1.6) AT 4CM DEPLOYMENT. PROXIMAL LEFT URETERAL THERMAL INJURY WITH URINE LEAK *DESPITE >1CM SAFETY MARGIN OF DEVICE FROM THE URETER PT REQUIRED PERCUTANEOUS DRAINAGE, URETER STENT PLACEMENT, REPEAT HOSPITALIZATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARBURST TALON SEMI-FLEX | RFA PROBE | GEI | ANGIODYNAMICS | NA | 551686 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Other |