UNICEL® DXC 800 SYNCHRON® SYSTEM
Report
- Report Number
- 2050012-2012-00163
- Event Type
- Malfunction
- Date Received
- January 23, 2012
- Date of Event
- December 30, 2011
- Report Date
- December 30, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JGJ
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BECKMAN COULTER INC. EVALUATED THE INSTRUMENT BY LEADING THE CUSTOMER THOUGH ROUTINE TROUBLESHOOTING ACTIVITIES. CUSTOMER TECHNICAL SERVICES RECOMMENDED THAT THE CUSTOMER FLUSH THE CHEMISTRY CARTRIDGE (CC) SAMPLE PROBE, AND THEN RECALIBRATE AND PERFORM A PRECISION RUN ON CHEMISTRY QUALITY CONTROL (QC) RESULTS. FOLLOW-UP WITH THE CUSTOMER INDICATED THAT THE CUSTOMER HAD CLEANED THE CC PROBE AND HAD EXECUTED A TEN REPETITION QC PRECISION ASSESSMENT WITH ACCEPTABLE RESULTS. THE MG CARTRIDGE WAS NO AVAILABLE FOR BECKMAN COULTER INC. ASSESSMENT. ROOT CAUSE IS NOT KNOWN BUT BECKMAN COULTER INC. LED CUSTOMER TROUBLESHOOTING APPEARS TO HAVE RESOLVED THE ISSUE.
A CUSTOMER REPORTED THAT ON (B)(6) 2011 ERRONEOUS, ELEVATED MAGNESIUM (MG) RESULTS WERE GENERATED ON A UNICEL DXC 800 SYNCHRON SYSTEM FOR THREE PATIENT SAMPLES. BECKMAN COULTER INC. ASSESSMENT OF CUSTOMER SUPPLIED PATIENT DATA INDICATED THAT INITIAL AND CRITICAL RERUN RESULTS WERE ELEVATED WHEN COMPARED TO REPEAT RESULTS SUBSEQUENTLY GENERATED. THE REPEAT RESULTS WERE ASSOCIATED WITH THE USE OF A NEW REAGENT PACK ACCORDING TO THE CUSTOMER. OTHER CHEMISTRIES WERE RUN FOR THESE SAMPLES BUT WERE NOT REPEATED. NO ERRONEOUS MG RESULTS WERE RELEASED FROM THE LABORATORY AND HENCE THERE WAS NO DEATH, SERIOUS INJURY OR MODIFICATION TO PATIENT TREATMENT ASSOCIATED OR ATTRIBUTED TO THIS EVENT. THE INSTRUMENT MG QUALITY CONTROL (QC) RESULTS DURING THE TIME FRAME OF THE EVENT RECOVERED WITHIN THE CUSTOMER'S ESTABLISHED SPECIFICATIONS. NO SAMPLE ISSUES WERE NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 800 SYNCHRON® SYSTEM | ANALYZER, CHEMISTRY | JGJ | BECKMAN COULTER, INC. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SYNCHRON MAGNESIUM REAGENT |