FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (HS-TNI ) CARTRIDGE

MDR report key: 24247120 · Received February 3, 2026

Report

Report Number
2245578-2026-00024
Event Type
Malfunction
Date Received
February 3, 2026
Date of Event
January 25, 2026
Report Date
March 23, 2026
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K240984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: (B)(4). THE INVESTIGATION WAS COMPLETED ON 18-MAR-2026. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AP (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR HS-TNI CARTRIDGE LOT B25338C.

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

NA.

Description of Event or Problem · 0

ON 01/30/-2026, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULTS ON A 62 YEAR OLD FEMALE PATIENT PRESENTED WITH DIZZINESS, ELEVATED TROPONINS, & ATYPICAL CHEST PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. METHOD DATE COLLECTED TESTED RESULTS (NG/L) SAMPLE POSITIVE/NEGATIVE I-STAT (B)(6) 2026 17:39 17:39 87 A POSITIVE. I-STAT (B)(6) 2026 19:54 19:54 103 B POSITIVE. ALINITY (B)(6) 2026 1:12 1:12 <3 C NEGATIVE. ALINITY (B)(6) 2026 3:23 3:23 <3 D NEGATIVE. POSITIVE CUT-OFF VALUE FOR I-STAT HS-TNI TEST: >13 NG/L POSITIVE CUT-OFF VALUE FOR COMPARATIVE INSTRUMENT: >14 PG/ML. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI SEX N (NG/L, PG/ML) (NG/L, PG/ML) FEMALE 490 13 (10, 17) MALE 404 28 (19, 58). OVERALL 895 21 (14, 30).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
172623 I-STAT TROPONIN (HS-TNI ) CARTRIDGE HS-TNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA B25338C

Patients

Seq Age Sex Outcome Treatment
1 62 YR Female