QUANTUM
Report
- Report Number
- 3006073153-2026-00014
- Event Type
- Malfunction
- Date Received
- February 3, 2026
- Date of Event
- January 23, 2026
- Report Date
- May 26, 2026
- Manufacturer
- SPECTRUM MEDICAL LTD
- Product Code
- DTX
- UDI-DI
- 05060434422064
- PMA / PMN Number
- K202733
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRELIMINARY RESULTS SHOWED THAT WITH NO LEVEL SENSOR CONNECTED, THE SYSTEM DISPLAYS LEVEL 2 ON QWS AND SPT. WHEN THE LEVEL SENSOR IS CONNECTED, THE SYSTEM DISPLAYS LEVEL 0, AND NO LIGHT IS EMITTED FROM THE SENSOR. THIS INDICATES A POTENTIAL FAULT WITH THE PCBA - FPL. ROOT CAUSE AND INVESTIGATION SUMMARY WITH FOLLOW IN FINAL REPORT.
THE DEVICE IS BEING RETURNED AND AN INVESTIGATION IS ONGOING.
PRELIMINARY RESULTS SHOWED THAT WITH NO LEVEL SENSOR CONNECTED, THE SYSTEM DISPLAYS LEVEL 2 ON QWS AND SPT. WHEN THE LEVEL SENSOR IS CONNECTED, THE SYSTEM DISPLAYS LEVEL 0, AND NO LIGHT IS EMITTED FROM THE SENSOR. THIS INDICATES A POTENTIAL FAULT WITH THE PCBA - FPL. THE DEVICE HAS NOW BEEN RETURNED AND THE ROOT CAUSE AND INVESTIGATION SUMMARY WILL FOLLOW IN A FINAL REPORT.
THE QUANTUM VENTILATION MODULE UNITS WERE FUNCTIONING AS EXPECTED; HOWEVER, WHEN THE LEVEL SENSOR WAS CONNECTED TO THE QVM, IT FAILED TO OPERATE PROPERLY. WHEN THE PROTECTED/SAFE LEVEL SAFETY WAS ENGAGED, THE LED ON THE LEVEL SENSOR DID NOT ILLUMINATE WHEN FLUID WAS PRESENT.
THE QUANTUM VENTILATION MODULE UNITS WERE FUNCTIONING AS EXPECTED; HOWEVER, WHEN THE LEVEL SENSOR WAS CONNECTED TO THE QVM, IT FAILED TO OPERATE PROPERLY. WHEN THE PROTECTED/SAFE LEVEL SAFETY WAS ENGAGED, THE LED ON THE LEVEL SENSOR DID NOT ILLUMINATE WHEN FLUID WAS PRESENT.
THE QUANTUM VENTILATION MODULE UNITS WERE FUNCTIONING AS EXPECTED; HOWEVER, WHEN THE LEVEL SENSOR WAS CONNECTED TO THE QVM, IT FAILED TO OPERATE PROPERLY. WHEN THE PROTECTED/SAFE LEVEL SAFETY WAS ENGAGED, THE LED ON THE LEVEL SENSOR DID NOT ILLUMINATE WHEN FLUID WAS PRESENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 17025 | QUANTUM | QUANTUM VENTILATION MODULE | DTX | SPECTRUM MEDICAL LTD | QVM2-N-HM | 05060434422064 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |