FDA Adverse Event Malfunction Summary report: N

STATSTRIP GLUCOSE HOSPITAL METER SYSTEM

MDR report key: 24222320 · Received January 30, 2026

Report

Report Number
1219029-2026-00008
Event Type
Malfunction
Date Received
January 30, 2026
Date of Event
January 3, 2026
Report Date
May 20, 2026
Manufacturer
NOVA BIOMEDICAL CORP.
Product Code
PZI
PMA / PMN Number
K232075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION BEGUN BUT NOT YET COMPLETED. A SUPPLEMENTAL REPORT WILL BE FILED UPON COMPLETION OF INVESTIGATION.

Description of Event or Problem · 0

STATSTRIP METER SENDING A BLANK NTE SEGMENT TO CUSTOMER'S CONNECTIVITY SOFTWARE (RALS). THIS CAUSED THE RESULT TO NOT TRANSMIT TO RALS AND PREVENTED OTHER RESULTS BEHIND THIS ONE TO ALSO NOT TRANSFER TO THE PATIENTS CHART. ALTHOUGH NO SPECIFIC ALLEGATIONS WERE MADE REGARDING DELAYS IN TREATMENT, THIS ISSUE COULD HAVE LED TO POTENTIAL DELAYS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
85781 STATSTRIP GLUCOSE HOSPITAL METER SYSTEM PRESCRIPTION USE BLOOD GLUCOSE METER FOR NEAR-PATIENT TESTING PZI NOVA BIOMEDICAL CORP. 2.0

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown