FDA Adverse Event
Malfunction
Summary report: N
STATSTRIP GLUCOSE HOSPITAL METER SYSTEM
MDR report key: 24222320
·
Received January 30, 2026
Report
- Report Number
- 1219029-2026-00008
- Event Type
- Malfunction
- Date Received
- January 30, 2026
- Date of Event
- January 3, 2026
- Report Date
- May 20, 2026
- Manufacturer
- NOVA BIOMEDICAL CORP.
- Product Code
- PZI
- PMA / PMN Number
- K232075
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
INVESTIGATION BEGUN BUT NOT YET COMPLETED. A SUPPLEMENTAL REPORT WILL BE FILED UPON COMPLETION OF INVESTIGATION.
Description of Event or Problem · 0
STATSTRIP METER SENDING A BLANK NTE SEGMENT TO CUSTOMER'S CONNECTIVITY SOFTWARE (RALS). THIS CAUSED THE RESULT TO NOT TRANSMIT TO RALS AND PREVENTED OTHER RESULTS BEHIND THIS ONE TO ALSO NOT TRANSFER TO THE PATIENTS CHART. ALTHOUGH NO SPECIFIC ALLEGATIONS WERE MADE REGARDING DELAYS IN TREATMENT, THIS ISSUE COULD HAVE LED TO POTENTIAL DELAYS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 85781 | STATSTRIP GLUCOSE HOSPITAL METER SYSTEM | PRESCRIPTION USE BLOOD GLUCOSE METER FOR NEAR-PATIENT TESTING | PZI | NOVA BIOMEDICAL CORP. | 2.0 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |