FDA Adverse Event Injury Summary report: N

INSET II

MDR report key: 24219860 · Received January 30, 2026

Report

Report Number
3003442380-2026-00212
Event Type
Injury
Date Received
January 30, 2026
Date of Event
January 2, 2026
Report Date
February 5, 2026
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244018129
PMA / PMN Number
K032854
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INITIAL AND FINAL (B)(4) - DEVICE 2 OF 3.

Additional Manufacturer Narrative · 0

SUPPLEMENTAL REPORT 01- (B)(4) - MDR 3003442380-2026-00212. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH: A QUERY WAS RUN IN THE EQMS ON 05-FEB-2026 AGAINST "LOT NUMBER 6010351 AND SIMILAR MALFUNCTION CODES: SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, THE REVIEW CONFIRMED THAT LOT 6010351 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NCRS OR CORRECTIVE AND PREVENTIVE ACTION (CAPA)S OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 05-FEB-2026 AGAINST "LOT NUMBER" CRITERIA EQUAL 6010351 AND SIMILAR MALFUNCTION CODES: SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE. THE COUNT OF COMPLAINT IS 4. THE COMPLAINT NUMBERS ARE COMPLAINT (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010351 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN THE LINEA 04 ON 24-NOV-2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. TEST RESULTS: IN ORDER TO TEST THE PRODUCT, NO PHOTO OR PHYSICAL SAMPLE WAS PROVIDED, THEREFORE, THE REFERENCE SAMPLES FROM THE LOT HAVE BEEN REQUESTED. VISUAL TEST ACCORDING TO WI VERSION 3 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST. FUNCTIONAL TEST AIR FLOW ACCORDING TO WI VERSION 2 ON REFERENCE SAMPLES, 10 SAMPLES OUT 10 SAMPLES PASSED THE TEST, FOR THE SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE. ALL TEST RESULTS WERE WITHIN SPECIFICATIONS AS PER THE TEST REPORT (B)(4) ATTACHED IN THIS RECORD. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO PHYSICAL SAMPLES WERE RETURNED. NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO THRESHOLDS IDENTIFIED FOR THE LOT IN QUESTION AND SERIOUS INJURY/DEATH, NO FURTHER ACTIONS ARE REQUIRED. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CAPA 1768101 AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS RESULT OF TRENDING ACTIVITIES. THIS IS A COMPLAINT WHICH IS BEING ADDRESSED AS PART OF A CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES WHICH HAS BEEN OPENED AS A RESULT OF TRENDING ACTIVITIES. THIS COMPLAINT IS A REPORTABLE WITH VALID LOT NUMBER/PRODUCT CODE. CAPA DETERMINATION RESULTS: THIS COMPLAINT FALLS UNDER THE SCOPE OF THE CAPA 1768101: "AUTOSOFT 90, KINKED SOFT CANNULA ISSUES" AND WILL COVER INSET I (AUTOSOFT XC DHF-13.8 & 13.13) AND INSET II (AUTOSOFT 90 DHF-14 AND 14.3) PRODUCTS. ROOT CAUSE OF PROBLEM: THE ROOT CAUSE HAS BEEN IDENTIFIED AS: METHOD, MANPOWER, MEASUREMENT, MACHINE: CORRECTIVE ACTION AS A RESULT OF THE INVESTIGATION: THE ACTION PLAN AND DEADLINES IS AS FOLLOWED: THE FOLLOWING ACTIONS WERE ALREADY IMPLEMENTED IN THE NC 1515780. 1. INCLUDE THE DEFECT IN DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE). 2. IMPROVE GUARDS OF THE CONVEYORS. 3. UPDATE TRAYS FOR THE STOCK OF CANNULAS. 4. UPDATE DOCUMENT 3A02003 (QUALITY SPECIFICATION FOR CATHETER FIXTURE FOR SKEWED CATHETERS) FOR INCLUDE THE HANDLING OF THE CATHETER DURING THE PROCESS. 5. UPDATE THE DOCUMENT 4805074 (WORK INSTRUCTION INSET LINE) TO INCLUDE THE PREVENTIVE ACTIONS IMPLEMENTED IN THE PROCESS (TRAYS). A REMAINING ACTION (TO ADDRESS DESIGN ROOT CAUSE) AND DEADLINES IS AS FOLLOWED: ADD CYLINDERS TO THE LID TO MAINTAIN IN POSITION THE NEEDLE AND CANNULA DURING TRANSPORTATION AND PRIOR USE (DATABASE (B)(4) - 30-SEP-2025). CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6010351 AND RELATED MALFUNCTION CODES. FOUR COMPLAINT WAS IDENTIFIED FOR THIS LOT; HOWEVER, NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. SAMPLES WERE REQUESTED; HOWEVER, THE CUSTOMER CONFIRMED THAT NO SAMPLES WERE AVAILABLE FOR ANALYSIS. CONSEQUENTLY, AN INVESTIGATION WAS CONDUCTED USING REFERENCE SAMPLES, AND NO FAILURES RELATED TO THE COMPLAINT WERE IDENTIFIED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION AND REFERENCE SAMPLE ANALYSIS ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN AUSTRALIA. IT WAS REPORTED THAT PATIENT FACED SIX INFUSION SET KINKED CANNULA EVENT ON (B)(6) 2026. THE PATIENT WENT TO EMERGENCY ROOM AND EVENTUALLY GOT HOSPITALIZED DUE TO HYPERGLYCEMIA EVENT. THE INSERTION SITE WAS ABDOMEN. THE PATIENT HAD HIGH KETONES AT THE TIME OF THE EVENT. THE BLOOD GLUCOSE LEVEL WAS 22 MMOL/L AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS FLUIDS OF SALINE AND INSULIN. THE PATIENT WAS RELEASED FROM THE HOSPITAL ON (B)(6) 2026.THE ISSUE OCCURED WITH 3 INFUSION SETS. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
286502 INSET II UNO INSET II 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1002817 6010351 05705244018129

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Hospitalization