OFP FLUSHING PUMP
Report
- Report Number
- 9611174-2026-00030
- Event Type
- Malfunction
- Date Received
- January 28, 2026
- Date of Event
- January 5, 2026
- Report Date
- January 28, 2026
- Manufacturer
- KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD.
- Product Code
- FEQ
- UDI-DI
- 15019778003528
- PMA / PMN Number
- K100899
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE THE RESULTS OF THE FINAL INVESTIGATION. UPDATED FIELDS: H2, H3, H6, H11. THE DEVICE WAS NOT RETURNED TO OLYMPUS FOR INSPECTION. HOWEVER, TECHNICAL ASSISTANCE CENTER (TAC) PROVIDED REMOTE TROUBLESHOOTING AND LOOSE WATER BOTTLE CAP WILL NOT PRODUCE AN AIRTIGHT SEAL AND WILL PRODUCE DRIBBLING FROM THE DISTAL END OF THE SCOPE AND THE REPORTED FAILURE WAS CONFIRMED. IN ADDITION, THE FOLLOWING REPORTABLE MALFUNCTIONS WERE IDENTIFIED DURING THE DEVICE EVALUATION: LOOSE WATER BOTTLE CAP. BASED ON THE RESULTS OF THE INVESTIGATION, DUE TO RETIGHTENED THE SEAL AND THE DRIBBLING OF WATER FROM THE DISTAL TIP OF THE SCOPE STOPPED IMMEDIATELY. A DEFINITIVE ROOT CAUSE COULD NOT BE IDENTIFIED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
THE INVESTIGATION IS ONGOING. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETED OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. TECHNICAL ASSISTANCE CENTER (TAC) PROVIDED EDUCATION VIA PHONE CALL CONFERENCE WITH THE CUSTOMER. TAC EDUCATED THE CUSTOMER THAT A LOSE OLYMPUS FLUSHING PUMP (OFP) WATER BOTTLE CAP WILL NOT PRODUCE AN AIRTIGHT SEAL AND WILL PRODUCE DRIBBLING FROM THE DISTAL END OF THE SCOPE. THE CUSTOMER STATED THE STAFF RETIGHTENED THE SEAL AND THE DRIBBLING OF WATER FROM THE DISTAL TIP OF THE SCOPE STOPPED IMMEDIATELY. TAC INFORMED THE CUSTOMER THAT THE PUMP HEAD ITEM #7502419 SHOULD BE CHANGED EVERY 12 MONTHS. TAC EMAILED THE CUSTOMER THE FOLLOWING STATEMENT: ¿THE ITEM LISTED BELOW SHOULD BE CHANGED EVERY 12 MONTHS. 7502419 (PUMP HEAD ASSEMBLY SPARE). THIS ITEM CAN BE ORDERED THROUGH OUR CUSTOMER CARE DEPT. BY CALLING 800-848-9024, OPTION 2.¿ THE ISSUE WAS RESOLVED. NO FURTHER TAC ASSISTANCE IS NEEDED.
NO ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER.
IT WAS REPORTED THE OLYMPUS FLUSHING PUMP HAD DRIBBLING FROM THE DISTAL END DURING A DIAGNOSTIC COLONOSCOPY. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 265087 | OFP FLUSHING PUMP | OLYMPUS FLUSHING PUMP | FEQ | KEYMED (MEDICAL AND INDUSTRIAL EQUIPMENT) LTD. | 7501349 | 15019778003528 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |