PUMP 1886 780G OUS BLE PUMP MG/DL
Report
- Report Number
- 2032227-2026-115659
- Event Type
- Malfunction
- Date Received
- January 28, 2026
- Date of Event
- January 16, 2026
- Report Date
- May 5, 2026
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZP
- UDI-DI
- 000076300063923501
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
UNIT RECEIVED PUMP ERROR 38 DURING REWIND. UNABLE TO PERFORM THE DISPLACEMENT, REWIND, PRIME/SEATING, BASIC OCCLUSION, FORCE SENSOR, OCCLUSION TESTS, AND SELF TESTS DUE TO THE PUMP ERROR 38. UNIT WAS SUCCESSFULLY DOWNLOADED USING THUMP SOFTWARE. ON ADAPT, NO RELEVANT ALARMS WERE NOTED. HOWEVER ON ADAPT, CONFIRMED (38) ALARM ON 01/16/2026 23:21:47.000 HOUR FOR ALARMS THAT MAY HAVE OCCURRED IN THE PAST AND LINE NUMBER 752 FILE NUMBER 121 01/16/2026 23:21:39.000 OR DURING A COMPLAINT CALL THAT MIGHT HAVE TRIGGERED THE REASON COMPLAINT. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND NO COMPONENT OR MOISTURE DAMAGE ON THE ELECTRONIC ASSEMBLY, BATTERY CONNECTOR, BATTERY TUBE, MOTOR, VIBRATOR DURING VISUAL INSPECTION. MOTOR FAILED MOTOR TEST DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. THE SC1 CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE FOLLOWING WERE NOTED DURING THE PHYSICAL INSPECTION: SCRATCHED CASE AND PILLOWING KEYPAD OVERLAY. PUMP ERROR 38 DURING REWIND AND IN THE PUMP HISTORY CONFIRMED DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
SELECT PATIENT INFORMATION CANNOT BE PROVIDED DUE TO REGIONAL PRIVACY REGULATIONS. THE REPORTED DEVICE IS NOT MARKETED IN THE UNITED STATES, BUT IT IS A SAME/SIMILAR DEVICE TO ONE THAT IS MARKETED INSIDE THE UNITED STATES. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
IT WAS REPORTED TO MEDTRONIC MINIMED THAT THE CUSTOMER RECEIVED PUMP ERROR 38 (NO MOTOR MOVEMENT OR NEGLIGIBLE MOVEMENT. RETRIES TO FREE A STUCK MOTOR FAILED) ALARM. THE CUSTOMER REPORTED NO ADVERSE EVENT. THE EVENT INVOLVED PRODUCT(S) MMT-1886. TROUBLESHOOTING WAS PERFORMED FOR PUMP ERROR ALARM. CUSTOMER WAS ABLE TO CLEAR THE ERROR BUT WAS UNABLE TO COMPLETE THE REWIND PROCESS. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE, AND THE PUMP WILL BE REPLACED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. MMT-1886 WAS REQUESTED, AND CUSTOMER RESPONSE WAS THE DEVICE WILL BE RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469998 | PUMP 1886 780G OUS BLE PUMP MG/DL | AUTOMATED INSULIN DOSING DEVICE SYSTEM, SINGLE HORMONAL CONTROL | OZP | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-1886 | NG4020582H | 000076300063923501 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |