FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 24201292 · Received January 28, 2026

Report

Report Number
9617229-2026-01819
Event Type
Injury
Date Received
January 28, 2026
Date of Event
October 23, 2025
Report Date
May 5, 2026
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON APRIL 06, 2026, WITH LOT NUMBER 3133663. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS RUPTURE: OBSERVED AN OPENING THROUGH MICROSCOPIC INSPECTION ASSESSED AS FOLD FLAW OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASE, WEAR ABRASION AND NON-PENETRATING NICK.) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322642 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3133663

Patients

Seq Age Sex Outcome Treatment
1