INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2026-01819
- Event Type
- Injury
- Date Received
- January 28, 2026
- Date of Event
- October 23, 2025
- Report Date
- May 5, 2026
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON APRIL 06, 2026, WITH LOT NUMBER 3133663. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS RUPTURE: OBSERVED AN OPENING THROUGH MICROSCOPIC INSPECTION ASSESSED AS FOLD FLAW OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASE, WEAR ABRASION AND NON-PENETRATING NICK.) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.
HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322642 | INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 3133663 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |