FDA Adverse Event Malfunction Summary report: N

CARDIOMEMS POWER SUPPLY

MDR report key: 24198470 · Received January 28, 2026

Report

Report Number
3004936110-2026-00182
Event Type
Malfunction
Date Received
January 28, 2026
Date of Event
January 26, 2026
Report Date
February 23, 2026
Manufacturer
ABBOTT MEDICAL
Product Code
MOM
PMA / PMN Number
P100045
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.

Additional Manufacturer Narrative · 0

ONE POWER SUPPLY AND ONE POWER CORD MODEL WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION VERIFIED THE POWER SUPPLY WAS FRAYED AND WIRES WERE EXPOSED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT.

Description of Event or Problem · 0

THE PATIENT REPORTED EXPOSED AND FRAYED WIRES OF THE POWER SUPPLY. THE CORDS WERE REPLACED AND THERE WERE NO ADVERSE CONSEQUENCES REPORTED BY THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
260367 CARDIOMEMS POWER SUPPLY SYSTEM, HEMODYNAMIC, IMPLANTABLE MOM ABBOTT MEDICAL CM1110

Patients

Seq Age Sex Outcome Treatment
1 67 YR Male