OPTILENE 8/0 (0,4) 60CM 2XDR6 CV2 RCP
Report
- Report Number
- 3003639970-2026-00053
- Event Type
- Malfunction
- Date Received
- January 28, 2026
- Report Date
- May 13, 2026
- Manufacturer
- B. BRAUN SURGICAL, S.A.
- Product Code
- GAW
- PMA / PMN Number
- SEE H11
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
G4: REPORTED DEVICE NOT MARKETED IN THE U.S.; HOWEVER, SIMILAR DEVICES OR DEVICES THAT SHARE COMPONENTS, RAW MATERIALS, PROCESS METHODS OR OTHER TECHNOLOGICAL CHARACTERISTICS ARE REGISTERED WITHIN THE U.S. UNDER PMA/510(K) NUMBER: K133890. IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.
IT WAS REPORTED AN ISSUE WITH OPTILENE SUTURE. THE CLIENT REPORTED THE FOLLOWING: ONLY THE NEEDLE BROKE; ONE THREAD BROKE, AND ANOTHER THREAD DETACHED FROM THE NEEDLE AS SOON AS IT WAS REMOVED FROM THE RACE-PACK. THE JUNCTION BETWEEN THE THREAD AND THE NEEDLE IS NOT SMOOTH BUT HAS BURRS, MAKING IT DIFFICULT TO PULL THE NEEDLE DURING STITCHING AND CAUSING THE THREAD TO FRAY/SPLIT. THE THREAD SPLIT INTO TWO STRANDS. PACKAGING ERROR: THE THREAD IS CAUGHT IN THE SEALING OF OUTER PACKAGE, AND WAS NOT SECURED IN THE RACEPACK (INNER PACKAGE). ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 469564 | OPTILENE 8/0 (0,4) 60CM 2XDR6 CV2 RCP | CARDIAC SUTURE | GAW | B. BRAUN SURGICAL, S.A. | C3097712 | 124471 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |