FDA Adverse Event Malfunction Summary report: N

CAPTURA SERRATED FORCEPS WITH SPIKE

MDR report key: 2419164 · Received December 21, 2011

Report

Report Number
1037905-2011-00796
Event Type
Malfunction
Date Received
December 21, 2011
Report Date
November 30, 2011
Manufacturer
WILSON-COOK MEDICAL, INC.
Product Code
FCL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION EVALUATION: A PRODUCT EVALUATION WAS NOT PERFORMED IN RESPONSE TO THIS REPORT BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT PROVIDED TO COOK FOR EVAL. THE REPORT COULD NOT BE CONFIRMED. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS NOT OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. INVESTIGATION CONCLUSIONS: WE COULD NOT CONDUCT A COMPLETE INVESTIGATION BECAUSE THE PRODUCT SAID TO BE INVOLVED WAS NOT RETURNED FOR EVAL. A DEFINITIVE CAUSE FOR THE REPORTED OBSERVATION COULD NOT BE DETERMINED. PRIOR TO DISTRIBUTION, ALL CAPTURA SERRATED FORCEPS ARE SUBJECTED TO A VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: A REVIEW OF THE COMPLAINT HISTORY WAS CONDUCTED. THE LIKELIHOOD OF OCCURRENCE IS CONSIDERED RARE. CORRECTIVE ACTION IS NOT WARRANTED AT THIS TIME BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS AND REASSESS THE RISK ASSESSMENT RESULTS AS POST MARKET FEEDBACK CONTINUES TO BECOME AVAILABLE.

Description of Event or Problem · 1

THE FOLLOWING INFO WAS PROVIDED BY THE COOK AREA REPRESENTATIVE BASED ON COMMENTS MADE BY THE USER: THE PHYSICIAN WAS TAKING SAMPLES IN THE ESOPHAGUS AND STOMACH. THE FORCEPS WERE TEARING THE TISSUE INSTEAD OF TAKING A CLEAN BITE. MEDICAL INTERVENTION WAS NOT REQUIRED IN RESPONSE TO THE REPORTED TEARING OF THE TISSUE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADD'L PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPTURA SERRATED FORCEPS WITH SPIKE FCL, FORCEPS, BIOPSY, NON-ELECTRIC FCL WILSON-COOK MEDICAL, INC. W3043404

Patients

Seq Age Sex Outcome Treatment
1 27 YR OLYMPUS ENDOSCOPE (UNK TYPE)