FDA Adverse Event Malfunction Summary report: N

SYRINGE 50ML LL TIP 1ML

MDR report key: 24189335 · Received January 27, 2026

Report

Report Number
1911916-2026-00042
Event Type
Malfunction
Date Received
January 27, 2026
Date of Event
January 21, 2026
Report Date
February 3, 2026
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
00382903096534
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

(B)(4) FOLLOW UP FOR DEVICE EVALUATION. A SYRINGE WAS REPORTED TO CONTAIN BLACK SPOTS AT THE TIP. TO SUPPORT THE INVESTIGATION, ONE SAMPLE IN AN OPENED BLISTER PACKAGE AND TWO PHOTOGRAPHS WERE RECEIVED AND EVALUATED BY THE QUALITY TEAM. VISUAL INSPECTION OF THE RETURNED SAMPLE IDENTIFIED FOUR EMBEDDED DARK SPECKS AT THE BOTTOM OF THE SYRINGE BARREL. THE PHOTOGRAPHS PROVIDED CORRESPOND TO THE SAMPLE RECEIVED, AND NO ADDITIONAL DEFECTS OR IMPERFECTIONS WERE OBSERVED. THE EMBEDDED SPECKS ARE CONSISTENT WITH DEGRADED RESIN, WHICH CAN OCCUR DURING INJECTION MOLDING AT START UP OR INTERMITTENTLY AND MAY DISLODGE AND BECOME INCORPORATED INTO MOLDED COMPONENTS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR MATERIAL NUMBER 309653, LOT 5212999. THIS REVIEW DID NOT IDENTIFY ANY QUALITY ISSUES DURING PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION, AND NO RELATED QUALITY NOTIFICATIONS WERE FOUND. ALL IN PROCESS AND FINAL INSPECTIONS MET SPECIFICATION REQUIREMENTS, AND NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED FOR THIS LOT TO DATE. BASED ON THE INVESTIGATION AND THE RETURNED SAMPLE EVALUATION, THE CUSTOMER REPORTED CONDITION IS CONFIRMED.

Description of Event or Problem · 0

IT IS REPORTED FOREIGN MATTER. DESCRIPTION: MATERIAL: ON (B)(6) 2026, ONE BD 50 ML SYRINGE LUER-LOK TIP (REF 309653) CONTAINED BLACK SPOTS AT THE SYRINGE TIP. LOT 5212999 EXP 2030-07-31, NOT UTILIZED ON PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
251793 SYRINGE 50ML LL TIP 1ML SYRINGE, PISTON FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 5212999 00382903096534

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown