FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 2418847 · Received January 17, 2012

Report

Report Number
9611451-2012-00021
Event Type
Malfunction
Date Received
January 17, 2012
Date of Event
November 17, 2011
Report Date
November 18, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: NINE COMPLAINT MR290V CHAMBERS WERE RECEIVED, WITH THE FOLLOWING LOT NUMBERS: LOT 090303: 1, LOT 090427: 1, LOT 090924:1, LOT 100914: 1, LOT 110114: 1, LOT 110719: 2, LOT 110805: 1, LOT 110806: 1. THE CHAMBERS WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED SCRATCHES OR STRESS MARKS ON THE DOMES OF ALL NINE CHAMBERS. IN ADDITION, THREE OF THE CHAMBERS HAD DENTS AND SCRATCHES ON THEIR BASES. CONCLUSION: THE NATURE OF THE CRACKING, AND THE FACT THAT IT WAS FOUND BEFORE USE OF THE CHAMBERS, SUGGESTS THAT THE DAMAGE OCCURED DURING TRANSPORTATION OR STORAGE OF THE CHAMBERS. EVERY MR290 CHAMBER IS PRESSURE TESTED TO 200CMH20 FOLLOWING THE MANUFACTURING PROCESS TO CHECK FOR ANY LEAKS PRESENT IN THE CHAMBER DOME DUE TO CRACKS AND OTHER CAUSES. ANY CHAMBER WHICH FAILS THIS TEST IS REJECTED. IN ADDITION, THE PRESSURE TEST IS FOLLOWED BY A VISUAL INSPECTION OF EACH CHAMBER. THIS SUGGESTS THAT THE CRACKS OCCURRED POST PRODUCTION. OUR USER INSTRUCTIONS THAT ACCOMPANY THE MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT WHEN UNPACKING SOME UNUSED MR290 AUTOFEED HUMIDIFICATION CHAMBERS THEY OBSERVED WHITE MARKS AND SCRATCHES ON THE CHAMBER DOMES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 090303

Patients

Seq Age Sex Outcome Treatment
1