FDA Adverse Event Injury Summary report: N

TUFF IMPLANT

MDR report key: 24186789 · Received January 27, 2026

Report

Report Number
3011390931-2026-00032
Event Type
Injury
Date Received
January 27, 2026
Date of Event
December 18, 2025
Report Date
January 27, 2026
Manufacturer
NORIS MEDICAL LTD
Product Code
DZE
UDI-DI
07290108812107
PMA / PMN Number
K140440
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DHR WAS REVIEWED FOR LOT #9001462 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE INVESTIGATION FOR THIS EVENT HAS NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. OUR MANUFACTURING SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFORM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025. ON (B)(6) 2025, LOOSENING OF IMPLANT NOT RELATED TO BONE INGROWTH WAS VERIFIED. ACCORDING TO THE INFORMATION, THE PATIENT IS A SMOKER. AT THE EVENT, THE PATIENT EXPERIENCED: PAIN. OBSERVED DURING THE EVENT: MOBILITY/INFECTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
250325 TUFF IMPLANT TUFF DENTAL IMPLANT DZE NORIS MEDICAL LTD NM-F3711 9001462 07290108812107

Patients

Seq Age Sex Outcome Treatment
1 46 YR Unknown