FDA Adverse Event
Malfunction
Summary report: N
VARIAN CLINAC
MDR report key: 241812
·
Received September 20, 1999
Report
- Report Number
- MW1017223
- Event Type
- Malfunction
- Date Received
- September 20, 1999
- Report Date
- September 15, 1999
- Manufacturer
- VARIAN ASSOC, INC.
- Product Code
- JAI
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- NC, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
UPON ROTATION OF THE TREATMENT COUCH ON ITS BASE A CRACKING NOISE WAS HEARD EMITTED FROM THE BASE OF THE TREATMENT COUCH. INVESTIGATION REVEALED ALL BUT ONE OF THE BOLTS WHICH SECURED THE TREATMENT COUCH TO THE BASE HAD BECOME SHEARED OFF. THIS ALLOWED THE TREATMENT COUCH TO TIP DOWNWARD TOWARD THE GANTRY AND STAND WHEN THE TABLE WAS EXTENDED TOWARD THE ACCELERATOR WITH A PT IN PLACE. THE RESULT WOULD BE MISALIGNMENT OF THE PT AT THE ISOCENTER BY APPROX 2 CM (ESTIMATED). PT TREATMENTS WERE DISCONTINUED, AND REPAIR OF THE UNIT REQUIRED RIGGERS TO REMOVE THE COUCH FROM THE FLOOR AND BASE, REPLACE ALL THE SHEARED BOLTS AND STAND-OFFS, AND RE-ESTABLISH ISOCENTER. OPERATION APPEARS TO BE NORMAL SINCE THE TIME OF THAT REPAIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VARIAN CLINAC | TREATMENT COUCH | JAI | VARIAN ASSOC, INC. | 2100C/D | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | Other |