FDA Adverse Event Malfunction Summary report: N

VARIAN CLINAC

MDR report key: 241812 · Received September 20, 1999

Report

Report Number
MW1017223
Event Type
Malfunction
Date Received
September 20, 1999
Report Date
September 15, 1999
Manufacturer
VARIAN ASSOC, INC.
Product Code
JAI
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

UPON ROTATION OF THE TREATMENT COUCH ON ITS BASE A CRACKING NOISE WAS HEARD EMITTED FROM THE BASE OF THE TREATMENT COUCH. INVESTIGATION REVEALED ALL BUT ONE OF THE BOLTS WHICH SECURED THE TREATMENT COUCH TO THE BASE HAD BECOME SHEARED OFF. THIS ALLOWED THE TREATMENT COUCH TO TIP DOWNWARD TOWARD THE GANTRY AND STAND WHEN THE TABLE WAS EXTENDED TOWARD THE ACCELERATOR WITH A PT IN PLACE. THE RESULT WOULD BE MISALIGNMENT OF THE PT AT THE ISOCENTER BY APPROX 2 CM (ESTIMATED). PT TREATMENTS WERE DISCONTINUED, AND REPAIR OF THE UNIT REQUIRED RIGGERS TO REMOVE THE COUCH FROM THE FLOOR AND BASE, REPLACE ALL THE SHEARED BOLTS AND STAND-OFFS, AND RE-ESTABLISH ISOCENTER. OPERATION APPEARS TO BE NORMAL SINCE THE TIME OF THAT REPAIR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VARIAN CLINAC TREATMENT COUCH JAI VARIAN ASSOC, INC. 2100C/D NA

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Other