FDA Adverse Event Other Summary report: N

PRECISE PAIN RELIEVING PATCH

MDR report key: 2417645 · Received January 11, 2012

Report

Report Number
1054358-2012-00001
Event Type
Other
Date Received
January 11, 2012
Date of Event
December 14, 2011
Report Date
January 10, 2012
Manufacturer
KIRIBAI KOBAYASHI PHARMACEUTICAL CO LTD
Product Code
IMD
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS IS THE FIRST REPORT MEDIHEAT HAS RECEIVED, IN FOUR YEARS OF PRODUCING THE DIRECT SKIN PRODUCT, OF AN IGNITION (A PATCH CATCHING FIRE). THE PRODUCT INVOLVED HAS BEEN REQUESTED FROM THE PT FOR A FULL EVAL. WE HAVE CONDUCTED A FULL QUALITY REVIEW OF ALL RECORDS AND HAVE CONFIRMED THAT THIS LOT MEETS ALL PRODUCT SPECS, AND THAT THERE IS NO EVIDENCE OF ANY NON-CONFORMANCE OR ABNORMALITIES IN THE MANUFACTURE OF THIS LOT. WE HAVE CURRENTLY NOT RECEIVED THE WARMER AT ISSUE, DESPITE THE PT'S AGREEMENT TO SEND THE WARMER. DUE TO THE NATURE OF THIS DEVICE, WE FEEL IT IS VERY UNLIKELY AND HIGHLY IMPROBABLE THAT THIS DEVICE COULD IGNITE. WE WILL COMPLETE OUR INVESTIGATION IF WE RECEIVE THE WARMER FROM THE PT, AND DETERMINE THE FACTORS THAT COULD HAVE CAUSED THE ALLEGED IGNITION. IF WE DO NOT HEAR BACK FROM THE PT, WE WILL CLOSE THE INVESTIGATION AND NOTIFY THE FDA.

Description of Event or Problem · 1

THIS WAS A SPONTANEOUS REPORT FROM A FEMALE OF AN UNSPECIFIED AGE FROM THE UNITED STATES: (B)(4). THE PT'S HEIGHT, WEIGHT AND MEDICAL HISTORY ARE UNSPECIFIED. THE PRECISE PAIN RELIEVING PATCH WAS APPLIED TO THE PT FOR "4 - 5 HOURS LAYING ON THE SOFA." THE PT THEN REPORTED THE "PATCH BURNED IN ONE SPOT AND I REMOVED. PATCH WAS ACTUALLY ON FIRE." THE PT STATED THAT THE ONLY INJURY WAS A "VERY VERY SMALL BURN ON MY BACK." MEDIHEAT, INC RECEIVED THE INITIAL REPORT ON 12/15/2011, AND RECEIVED PT AND LOT INFO ON 12/22/2011. THE PT WAS CONTACTED ON 12/22/2011 TO HAVE THE PRODUCT RETURNED FOR EVAL AND TESTING. THE PT RESPONDED ON 12/25/2011, THANKING US FOR THE QUICK RESPONSE AND NOTIFYING US THAT SHE WOULD SEND THE PRODUCT BACK THE FOLLOWING WEEK. AT THIS TIME MEDIHEAT INC, HAS NOT RECEIVED THE PRODUCT WITH THE ALLEGED IGNITION TO BE ABLE TO DETERMINE THE VALIDITY OF THE ALLEGATION. WE WILL ATTEMPT TO CONTACT THE PT ONE MORE TIME FOR ADD'L INFO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PRECISE PAIN RELIEVING PATCH NONE IMD KIRIBAI KOBAYASHI PHARMACEUTICAL CO LTD BDC005

Patients

Seq Age Sex Outcome Treatment
1 Other