FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 2417511 · Received January 10, 2012

Report

Report Number
2245578-2012-00011
Event Type
Malfunction
Date Received
January 10, 2012
Date of Event
December 21, 2011
Report Date
December 22, 2011
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

ON (B)(4) 2011, ABBOTT PONT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT RESULT OF 0.22 ON A PT WITH LOT U11195. SAMPLE COLLECTION: (B)(6) 2011, 08:45. SAMPLE TESTING: (B)(6) 2011, 09:00 I-STAT CTNI LOT U11259 0.03; (B)(6) 2011, 09:10 I-STAT CTNI FOR U11195 0.22. CUSTOMER STATES THAT THE PHYSICIAN USED THE CTNI VALUE OF 0.03 FOR PT TREATMENT AS THERE WAS NO CARDIAC. BASED ON THE INFO AVAILABLE AT THE TIME, THERE WERE NO INJURIES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA U11195

Patients

Seq Age Sex Outcome Treatment
1 63 YR