INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Report
- Report Number
- 9611451-2012-00018
- Event Type
- Malfunction
- Date Received
- January 16, 2012
- Date of Event
- December 19, 2011
- Report Date
- December 19, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE RT265 INFANT EVAQUA2 BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K034026. METHOD: THE COMPLAINT RT265 INFANT BREATHING CIRCUITS ARE NOT EXPECTED TO BE RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR INVESTIGATION AS THEY ARE NOT AVAILABLE. OUR ANALYSIS IS ACCORDINGLY BASED ON THE EVENT DESCRIPTION AND ADDITIONAL INFORMATION PROVIDED BY THE HOSPITAL. IN OUR INVESTIGATION, A PERFORMANCE TEST WAS COMPLETED ON A KNOWN GOOD RT265 CIRCUIT. RESULT: PERFORMANCE TEST REVEALED THAT WHEN THE PORT CAP WAS FULLY INSERTED, THE CIRCUIT PERFORMED WITHIN SPECIFICATION. WHEN THE PORT CAP WAS PARTIALLY INSERTED OR DISCONNECTED (SIMULATING THE REPORTED MISSING PORT CAP), THE CIRCUIT PERFORMED OUTSIDE OF SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINT OF THIS TYPE FOR LOT NUMBER 110825. CONCLUSION: THE PERFORMANCE TEST CONFIRMED THAT THE BREATHING CIRCUIT WOULD FAIL THE PRESSURE TEST IF A PRESSURE PORT CAP IS NOT FULLY INSERTED OR MISSING FROM THE BREATHING CIRCUIT. ALL RT265 BREATHING CIRCUITS ARE PRESSURE AND FLOW TESTED DURING PRODUCTION. IT IS POSSIBLE THAT THE PORT CAP MAY HAVE BEEN DISCONNECTED AFTER THE PRESSURE TEST.
A HOSPITAL IN (B)(6) REPORTED VIA A FISHER & PAYKEL HEALTHCARE SALES REPRESENTATIVE THAT THE PORT OF THE EXPIRATORY LIMB OF FOUR RT265 INFANT BREATHING CIRCUIT KITS WAS MISSING. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT265 | 110825 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |