FDA Adverse Event Malfunction Summary report: N

BD BBL¿ GRAM SAFRANIN

MDR report key: 24168079 · Received January 23, 2026

Report

Report Number
1119779-2026-00024
Event Type
Malfunction
Date Received
January 23, 2026
Date of Event
January 7, 2026
Report Date
March 3, 2026
Manufacturer
BECTON DICKINSON & CO. (SPARKS)
Product Code
JTS
UDI-DI
30382902125311
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY - THIS MEMO IS TO SUMMARIZE FINDINGS ON YOUR COMPLAINT RELATED TO BOTTLE GRAM SAFRANIN 250ML, CATALOG NUMBER 212531, BATCH NUMBER 4242097, COMPLAINT # (B)(4), FOR UNSATISFACTORY PERFORMANCE. COMPONENTS ARE MIXED AND DISPENSED INTO THE APPROPRIATE CONTAINERS. AFTER QC TESTING PRODUCT IS RELEASED AND TRANSPORTED TO THE DISTRIBUTION CENTER. THE COMPLAINT INVESTIGATION INCLUDED A BATCH HISTORY REVIEW OF GRAM CRYSTAL VIOLET. THE BATCH HISTORY RECORD REVIEW INDICATED NO DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY. COMPLAINT TRENDS WERE REVIEWED FOR A PERIOD COVERING 12 MONTHS. DURING THAT TIME THERE HAVE BEEN NO CONFIRMED COMPLAINTS FOR THE DEFECT IN QUESTION FOR THIS PRODUCT. ONE PHOTO WAS RECEIVED. THE PHOTO DEPICTS A MICROSCOPIC VIEW OF GRAM-NEGATIVE CULTURE AND TWO DARK PURPLE NONBACTERIAL ARTIFACTS. RETURN WAS RECEIVED ON 02/26/26 AFTER TESTING WAS COMPLETED. PACKAGE WAS RECEIVED DAMAGED AND SPILLED WITH MIXED MATERIALS. CONTENT INSIDE NOW MIXED MATERIALS AND CAN BE CONTAMINATED. NO FURTHER TESTING CAN BE CONDUCTED THIS TIME A RETENTION SAMPLE FROM THE COMPLAINT BATCH OF GRAM SAFRANIN WAS EVALUATED ALONG WITH A CONTROL BATCH. STAINING WAS COMPLETED PER INSTRUCTIONS IN THE PRODUCT INSERT. SLIDES OF E. COLI ATCC 25922 AND S. AUREUS ATCC 25923 CONTROLS WERE EVALUATED AND HAD SATISFACTORY STAINING RESULTS FOR THE RETENTION AND CONTROL BATCHES. THE RETENTION SAMPLE WAS COMPARABLE TO THE CONTROL. THIS COMPLAINT CANNOT BE CONFIRMED. WATERS WILL CONTINUE TO TREND DCM COMPLAINTS FOR UNSATISFACTORY PERFORMANCE.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING ONE (1) BD BBL¿ GRAM SAFRANIN, THERE WERE ARTIFACTS VISIBLE IN THE SLIDE, RENDERING THE STAIN UNUSABLE. USER REPORTS THE APPEARANCE OF THE ARTIFACTS APPEARED SIMILAR TO GRAM POSITIVE COCCI. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING ONE (1) BD BBL¿ GRAM SAFRANIN, THERE WERE ARTIFACTS VISIBLE IN THE SLIDE, RENDERING THE STAIN UNUSABLE. USER REPORTS THE APPEARANCE OF THE ARTIFACTS APPEARED SIMILAR TO GRAM POSITIVE COCCI. THERE WAS NO HEALTH IMPACT OR CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
233585 BD BBL¿ GRAM SAFRANIN STAINS, MICROBIOLOGIC JTS BECTON DICKINSON & CO. (SPARKS) 4242097 30382902125311

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown