FDA Adverse Event Malfunction Summary report: N

QUICK SET

MDR report key: 24148720 · Received January 22, 2026

Report

Report Number
3003442380-2026-00093
Event Type
Malfunction
Date Received
January 22, 2026
Date of Event
January 6, 2026
Report Date
March 26, 2026
Manufacturer
UNOMEDICAL UM-D
Product Code
FPA
UDI-DI
05705244017450
PMA / PMN Number
K160648
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH: A QUERY WAS RUN IN THE EQMS ON 26/MAR/2026 AGAINST "LOT NUMBER" "6012918" AND SIMILAR MALFUNCTION CODES: PRIMARY PACK DEFECT- TORN, RIPPED, CONTAINS HOLES, PRIMARY PACK HAS INCOMPLETE OR OPEN SEAL/WELD, IS TORN, RIPPED, CONTAINS HOLES, EXHIBITS EXTERNAL CONTAMINATION (E.G., WATER MARKS, STAINS) OR PRODUCT IS TRAPPED IN PACKAGING (E.G., TRAPPED IN WELD). THE REVIEW CONFIRMED THAT LOT 6012918 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NONCONFORMING REPORTS (NCR)S OR CORRECTIVE AND PREVENTIVE ACTION (CAPA)S OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 26/MAR/2026 AGAINST "LOT NUMBER" CRITERIA EQUAL "6012918" AND SIMILAR MALFUNCTION CODES: PRIMARY PACK DEFECT- TORN, RIPPED, CONTAINS HOLES, PRIMARY PACK HAS INCOMPLETE OR OPEN SEAL/WELD, IS TORN, RIPPED, CONTAINS HOLES, EXHIBITS EXTERNAL CONTAMINATION (E.G., WATER MARKS, STAINS) OR PRODUCT IS TRAPPED IN PACKAGING (E.G., TRAPPED IN WELD). THE COUNT OF COMPLAINTS IS 1. THE COMPLAINTS NUMBERS ARE (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6012918 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND MANUFACTURED IN THE MACHINE MULTIVAC 12, ON 22/APR/2025 WITH A TOTAL OF (B)(4) UNITS. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED IN ACCORDANCE WITH APPLICABLE PROCEDURES. ALL REQUIRED IN-PROCESS AND FINAL TESTS WERE COMPLETED AND MET SPECIFIED REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED THAT RELATE TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW SUPPORTS COMPLIANCE WITH MANUFACTURING AND QUALITY REQUIREMENTS; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. CAPA DETERMINATION ASSESSMENT - CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI (MONTHLY TRIPS AND ALERTS). COMPLAINT UNCONFIRMED: FOLLOWING INVESTIGATION THE REPORTED FAILURE COULD NOT BE CONFIRMED FOR THIS COMPLAINT. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6012918 AND RELATED MALFUNCTION CODES FOR DETACHMENT. ONE COMPLAINT WAS IDENTIFIED FOR THIS LOT, BUT NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED, AND NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN PORTUGAL. IT WAS REPORTED THAT PATIENT FACED PACKING ISSUE ON (B)(6) 2026. THE PRODUCT BEING REPORTED AS DAMAGED UPON RECEIPT PRIOR TO USE. THE PATIENT REPLACED THE INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
214875 QUICK SET UNO QUICK-SET 60/6 SC1 MECA FPA UNOMEDICAL UM-D MMT-399A 6012918 05705244017450

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown